Device information record format – Fluke Biomedical 601PRO XL User Manual
Page 108

6 0 1 P R O S E R I E S
X L
6 - 6
⇒ The Test Standard field must be an exact match with IEC 601-1,
VDE 701, VDE 751.1, HEI 95, AAMI, IEC 1010, AS/NZS 3551,
or the stored custom program names (CUSTOM1, CUSTOM2,
CUSTOM3, CUSTOM4). An exact match includes spaces and
punctuation marks, and is case sensitive.
⇒ If the standard is VDE 701, valid Types are T or F.
⇒ If the standard is IEC 601, HEI 95, VDE 751, AS/NZS 3551, or
IEC 1010, valid Types are B, BF, or CF.
⇒ If the standard is AAMI, valid Class/Type combinations are I/B,
I/BF, I/CF or II/B.
⇒ Valid Lead AP designations are RA, RL, LA, LL, or V1-V6. If the
field is empty, it means that this Lead AP is not defined.
⇒ In Custom Standards, Class and Type are ignored; however, error
checking for a match with whatever is there still occurs.
⇒ Valid Leads are 0, 1, 2, 3, 4, or 5.
⇒ Valid Power Off Delays and Startup Delays are 0 through 999.
Leading zeros or spaces are ignored.
⇒ Applied Current must be 1, 10, or 25.
5. Device Information Record Format
Each field is surrounded by double quotes (“) and separated by
commas. A Carriage Return (0DH or Ctrl-M), but no comma
appears after the last field in each record. An ETX character (03H
or Ctrl-C) signals the end of the transmission by the 601PRO. An
example of the Device Information format (export and import) is
shown below. Fields in braces [] are present only for compatibility
with older versions of the 601PRO and are ignored in the 601PRO
series
XL
. The fields are described elsewhere in this chapter.
"control #", "Procedure ID","Location","Device
Type","Manufacturer","Serial #",”[compatibility-
nothing]”,”[compatibility-nothing]”,"Class","Type","Test
Standard",”[compatibility-nothing]”,"# Leads", "Lead1 AP”,"Lead1
Type","Lead2 AP","Lead2 Type","Lead3 AP","Lead3 Type","Lead4
AP","Lead4 Type","Lead5 AP","Lead5 Type","Power Off Delay","Startup
Delay","Applied Current"
Example:
"CN123456","Procedure23","Upstairs","DEFIB","HEWLETT","654987","
","","II","BF","IEC 601-
1","","4","RA","BF","RL","BF","LA","CF","LL","CF","","","5","4","25"<
CR>