Appendices, A. specifications, Environmental – Fluke Biomedical 601PRO XL User Manual

Page 165: True rms, Specifications

background image

A-1

Specifications

This appendix details the device specifications for the 601PRO Safety Analyzer.

For more information, please contact your Fluke Biomedical Service representative.

Environmental

Operating

Temp: 10

°C to 40° C (50°F to 104° F)

Storage

Temp:

-20

° to +50° C (-4°F to +122° F)

Humidity:

80%

Non-Condensing

Operating Voltage Range

for full performance: 90-132 V RMS

198-264 V RMS

47-63 Hz

≤ 15 A

Front Panel Receptacle:

15 Amps @ 120 V

15 Amps 1/2 HP Max @ 250 V

True RMS

IEC-601/AAMI

IEC-1010

Measurement Range:

0.0-8000 µA True RMS
(per IEC 601-1 filter)

0.0-16000 µA True RMS**

Accuracy:

DC -1 kHz ± 1% of
reading ± 1 µA

DC -1 kHz ± 2% of
reading ± 2 µA

1 kHz-100 kHz ± 2% of
reading ± 1µA

1 kHz -100 kHz ± 4% of
reading ± 2 µA

100 kHz -1 MHz ± 5% of
reading ± 1µA

100 kHz -1 MHz ± 10%
of reading ± 10 µA

Input Impedance:

Per selected test load
specification

Per IEC-1010 specification

Frequency Response:

DC and 25 Hz -1 MHz
per selected test load
specification*

DC and 25 Hz -1 MHz
per selected test load
specification*

*Accuracy at frequencies below approximately 25 Hz (but not DC) is not specified
(because of "ripple" effects).

**For some leakage tests, the displayed range may be 0.0-9999 when using the IEC
1010 test load.