Appendices, A. specifications, Environmental – Fluke Biomedical 601PRO XL User Manual
Page 165: True rms, Specifications

A-1
Specifications
This appendix details the device specifications for the 601PRO Safety Analyzer.
For more information, please contact your Fluke Biomedical Service representative.
Environmental
Operating
Temp: 10
°C to 40° C (50°F to 104° F)
Storage
Temp:
-20
° to +50° C (-4°F to +122° F)
Humidity:
80%
Non-Condensing
Operating Voltage Range
for full performance: 90-132 V RMS
198-264 V RMS
47-63 Hz
≤ 15 A
Front Panel Receptacle:
15 Amps @ 120 V
15 Amps 1/2 HP Max @ 250 V
True RMS
IEC-601/AAMI
IEC-1010
Measurement Range:
0.0-8000 µA True RMS
(per IEC 601-1 filter)
0.0-16000 µA True RMS**
Accuracy:
DC -1 kHz ± 1% of
reading ± 1 µA
DC -1 kHz ± 2% of
reading ± 2 µA
1 kHz-100 kHz ± 2% of
reading ± 1µA
1 kHz -100 kHz ± 4% of
reading ± 2 µA
100 kHz -1 MHz ± 5% of
reading ± 1µA
100 kHz -1 MHz ± 10%
of reading ± 10 µA
Input Impedance:
Per selected test load
specification
Per IEC-1010 specification
Frequency Response:
DC and 25 Hz -1 MHz
per selected test load
specification*
DC and 25 Hz -1 MHz
per selected test load
specification*
*Accuracy at frequencies below approximately 25 Hz (but not DC) is not specified
(because of "ripple" effects).
**For some leakage tests, the displayed range may be 0.0-9999 when using the IEC
1010 test load.
Appendix
A