This chapter provides an overview of 601PRO

Auto

and

Step

mode sequences. The

Auto

or

Step

sequence performs all electrical safety tests required based on the Standard,

Class, and Type designated at the start of the test. Chapter 8, Standards and Principles,

provides a complete list and the order of execution of tests, which are detailed in Section

3: Refer to Test Limits for the Selected Standard. The operator may either recall a Device

Record based on its control number if Device Records were selected at setup, or a

Template based on its name if Templates were selected at setup, or use the instrument’s

present settings to execute a sequence.

Note:

The 601PRO allows the operator to make multiple measurements for the

enclosure leakage test while executing an

Auto

sequence. This allows the

operator to probe the DUT and find the highest reading before

continuing. Chapter 2, Setting Up the 601PRO, provides information for

enabling this feature in Section 7: Configure Enclosure Leakage for Auto

Mode Sequence.

1. Selecting Auto or Step Mode Testing
2. Executing Auto and Step Mode Tests
3. Creating/Editing a Device Record or Template

1. Selecting Auto or Step Mode Testing

• From the 601PRO

MAIN MENU

, press

TESTS/AUTOMODES

:

baw159f.eps

• Select either

AUTO

mode or

STEP

mode:

baw068f.eps

Note:

The following instructions and displays are based on having

device

records

selected during setup. If

templates

was selected during setup,

then replace

control number

with

template name

in the following

displays examples.