I N T R O D U C T I O N A N D D E S C R I P T I O N

1 - 7

Keys Used to Enter Device Control Numbers

The keys described below (0 through

) can be used to enter test control

numbers in the auto/step modes of operation. These keys are also referred to as

test-shortcut keys

and can be used to initiate manual tests.

0

Volts

In single lead mode, displays mains voltage.

In dual lead mode, displays voltage between

RED and BLACK test leads.

1 Current

Measures the current consumption (in

amperes) of the device under test.

2

Insulation

Tests insulation resistance (mains to case or

applied parts to case).

3 Protective

Earth

Resistance

Measures the earth resistance using a 1A test

current (unless 10A or 25A is selected).

4 Earth

Leakage Is measured between the DUT “Protective

Earth” terminal and the “Protective Earth”

terminal of the 601PRO.

5 Enclosure

Leakage

In single-lead mode, measures the enclosure

leakage (RED test lead to DUT protective

earth on the 601PRO).

6

Patient

Leakage

Measures the patient leakage current (applied

part to earth.

7

Mains on

Applied Part

Leakage

Applies 110% mains voltage to selected

applied part and measures leakage to earth in

both normal and reverse polarity. Does not

apply to patient auxiliary selections.

8 Patient

Auxiliary

Current

Measures the leakage and biasing current

between applied parts.

9 IEC

1010

Accessible

Voltage/Leakage

Selects the IEC 1010 Test Load. Measures

accessible voltage through the RED jack to

601PRO earth. Allows access to the

accessible leakage test.

/

VDE Equivalent

Device Leakage

Selects the IEC 601 test load. Applies 110%

of mains voltage between L1/L2 and earth.

VDE Equivalent

Patient Leakage

Selects the IEC 601 test load. Applies 110%

of mains voltage to the selected applied part

and measures the leakage.