Manual operation – Fluke Biomedical 601PRO XL User Manual
Page 53

M A N U A L M O D E
3 - 1 3
7. Manual Operation
The 601PRO can be used to perform individual tests manually (
Manual
mode)
without the need to perform an automated sequence (
Auto
mode).
☛ Note: In Chapter 8, Standards and Principles, a table of limits for each test
standard, Class, Type, and DUT Outlet configuration is available. In
Manual
mode, tests that are not part of the currently selected standard
will operate, but the limits will be displayed as
Inv
(invalid).
The operator may perform any of the available tests by selecting
MANUAL
from the
TESTS/AUTOMODES
option. The following will be displayed
:
baw045f.eps
Press the
*MORE
key to view all possible tests.
☛
Note:
The IEC-1010 Accessible Leakage Current and Accessible Voltage tests
appear in the menus only when the IEC-1010 test standard is the currently
selected standard. Similarly, the VDE Equivalent Device Leakage Current
and VDE Equivalent Patient Leakage Current appear in the menus only
when the appropriate VDE test standard is the currently selected standard.