Manual operation – Fluke Biomedical 601PRO XL User Manual

Page 53

background image

M A N U A L M O D E

3 - 1 3

7. Manual Operation

The 601PRO can be used to perform individual tests manually (

Manual

mode)

without the need to perform an automated sequence (

Auto

mode).

Note: In Chapter 8, Standards and Principles, a table of limits for each test

standard, Class, Type, and DUT Outlet configuration is available. In

Manual

mode, tests that are not part of the currently selected standard

will operate, but the limits will be displayed as

Inv

(invalid).

The operator may perform any of the available tests by selecting

MANUAL

from the

TESTS/AUTOMODES

option. The following will be displayed

:

baw045f.eps

Press the

*MORE

key to view all possible tests.

Note:

The IEC-1010 Accessible Leakage Current and Accessible Voltage tests

appear in the menus only when the IEC-1010 test standard is the currently

selected standard. Similarly, the VDE Equivalent Device Leakage Current

and VDE Equivalent Patient Leakage Current appear in the menus only

when the appropriate VDE test standard is the currently selected standard.