Vde 701 and vde 751-1, Automatic operation – Fluke Biomedical 601PRO XL User Manual
Page 118

6 0 1 P R O S E R I E S
X L
8 - 4
VDE 701 and VDE 751-1
VDE 701 and VDE 751 tests are specified for line voltages of >180V. Earth resistance test current is
set to 25 A.
☛
Note:
The Equivalent Patient Leakage test is only performed for VDE 751-1
.
VDE 701 Standard
DUT OUTLET
Class I
Class II
Description
Polarity
Circuit
T F T F
Mains
Voltage
V
N/A OFF NL NL NL NL
Earth Resistance
Ω
N/A
OFF
0.5
1
NT
NT
Equivalent Device Leakage µA
N/A
OFF
7000 7000 7000 7000
VDE 751.1 Standard
DUT OUTLET
Class I
Class II
Description
Polarity
Circuit B BF
CF B BF CF
Line Voltage V
N/A
OFF
NL
NL
NL
NL
NL
NL
Protective Earth
Resistance
Ω
N/A OFF
0.2
0.2
0.2
NT NT NT
Equivalent Device
Leakage µA
N/A OFF
1000
1000
1000
500 500 500
Equivalent Patient
Leakage µA
N/A OFF
NT 5000
50 NT 5000 50
Automatic Operation
There is no change in device screens or comments during
AUTO
or
STEP
mode operation with
VDE 751-1 and VDE 701. Test Limits are defined in the tables above.