beautypg.com

Computer control commands (cont.), Forces l2 closed to perform test – Fluke Biomedical 601PRO XL User Manual

Page 155

background image

C O M P U T E R C O N T R O L

1 0 - 7

Computer Control Commands (Cont.)

Auxiliary Current Measurement Commands (OUTLET AUTOMATICALLY ON FOR TEST)

COMMAND DESCRIPTION

NOTES

[ARA]

Aux. Leakage RA-ALL

601 Test Load

[ARL]

Aux. Leakage RL-ALL

601 Test Load

[ALA]

Aux. Leakage LA-ALL

601 Test Load

[ALL]

Aux. Leakage LL-ALL

601 Test Load

[AV1]

Aux. Leakage V1-V6-ALL

601 Test Load

[AARA]

Interlead Source Current RA-ALL

AAMI Test Load

[AARL]

Interlead Source Current RL-ALL

AAMI Test Load

[AALA]

Interlead Source Current LA-ALL

AAMI Test Load

[AALL]

Interlead Source Current LL-ALL

AAMI Test Load

[AAV1]

Interlead Source Current V1-V6-ALL

AAMI Test Load

Mains on Applied Part Commands

*

COMMAND DESCRIPTION

NOTES

[MAL]

Mains ALL-EARTH

601 Test Load

[MRA]

Mains RA-EARTH

601 Test Load

[MRL]

Mains RL-EARTH

601 Test Load

[MLA]

Mains LA-EARTH

601 Test Load

[MLL]

Mains LL-EARTH

601 Test Load

[MV1]

Mains V1-V6-EARTH

601 Test Load

[MCL]

Mains on AP CAL

601 Test Load

[RAL]

Rev Mains ALL-EARTH

601 Test Load

[RRA]

Rev Mains RA-EARTH

601 Test Load

[RRL]

Rev Mains RL-EARTH

601 Test Load

[RLA]

Rev Mains LA-EARTH

601 Test Load

[RLL]

Rev Mains LL-EARTH

601 Test Load

[RV1]

Rev Mains V1-V6-EARTH

601 Test Load

*Forces L2 closed to perform test.