Computer control commands (cont.) – Fluke Biomedical 601PRO XL User Manual

Page 154

background image

6 0 1 P R O S E R I E S

X L

1 0 - 6

Computer Control Commands (Cont.)

Leakage Current Measurement Commands (OUTLET AUTOMATICALLY ON FOR TEST)

COMMAND DESCRIPTION

NOTES

[ENC]

Enclosure No AP-GND

601 Test Load

[ENA]

Enclosure AP-GND

601 Test Load

[ERT]

*

Earth No AP-GND

601 Test Load

[ERA]

*

Earth AP-GND

601 Test Load

[DLL]

Dual Lead Leakage

[DLLD]

Dual Lead Leakage DC

[AERT]

*

Ground Wire Current No AP-GND

AAMI Test Load

[AERA]

*

Ground Wire Current AP-GND

AAMI Test Load

[AENC]

Chassis Source Current No AP-GND

AAMI Test Load

[AENA]

Chassis Source Current AP-GND

AAMI Test Load

*

Leaves earth connection open when executed.

Patient Leakage Measurement Commands (OUTLET AUTOMATICALLY ON FOR TEST)

COMMAND DESCRIPTION

NOTES

[PAL]

Pat. Leakage ALL-EARTH

601 Test Load

[PRA]

Pat. Leakage RA-EARTH

601 Test Load

[PRL]

Pat. Leakage RL-EARTH

601 Test Load

[PLA]

Pat. Leakage LA-EARTH

601 Test Load

[PLL]

Pat. Leakage LL-EARTH

601 Test Load

[PV1]

Pat. Leakage V1-V6-EARTH

601 Test Load

[APAL]

Patient Source Current ALL-GND

AAMI Test Load

[APRA]

Patient Source Current RA-GND

AAMI Test Load

[APRL]

Patient Source Current RL-GND

AAMI Test Load

[APLA]

Patient Source Current LA-GND

AAMI Test Load

[APLL]

Patient Source Current LL-GND

AAMI Test Load

[APV1]

Patient Source Current V1-V6-GND

AAMI Test Load