2. Errors and Suggested Corrective Actions

Follow the suggested corrective action for errors in the range of 0A00-1900. If the error

persists, you should collect as much information as possible about what you were doing at the

time of the error, and contact Fluke Biomedical or your local Service Center to report the

problem.

Code Description

Corrective

Action

0A00

System Test: 500-Volt
failure.

Recycle power and try again.

0B00

System Test: Mains
Voltage failure.

Recycle power and try again.

0C00

System Test: No Earth
connection.

Try a different wall outlet (see note below).

0D00

System Test Failure
preventing action.

The unit failed the power-up test(s). Recycle power and try again.

0E00

A to D read failure.

Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.

0F00

Calibration Value out of
range.

Confirm that patient leads are disconnected. If the problem
persists, contact Fluke Biomedical or your local Service Center to
report the problem.

1000

I Scale failure during
calibration.

Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.

1100

Voltage across 0.5 Ohm
cal resistor error.

Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.

1200 Calibration

Resistor

current error.

Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.

1300

System test leakage error.

Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.

1400

DIR checksum error.

Contact Fluke Biomedical or your local Service Center to report
the problem. The most probable cause is that Flash is worn out in
that sector.

1500

Result checksum error.

Contact Fluke Biomedical or your local Service Center to report
the problem. The most probable cause is that Flash is worn out in
that sector.

1600

No Test. Results storage
full. Delete old.

Delete some stored results to make room for new ones. Refer to
Chapter 5.

1700

Printer timeout. Print
buffer full.

You need to have a printer enabled and connected. Refer to
Chapter 2.

1800

Device Record Format is
invalid.

Review the format of the data being transmitted to the 601PRO.
Refer to Chapter 6.

1900

Serial Port timeout.
Cannot transmit.

Verify that you are using the correct serial cable, that it is properly
connected, and that the communications parameters are correctly
set-up on both the PC and the 601. Refer to Chapter 6.

2100

Cannot delete all Device
information until all test
results have been deleted.

Delete all test results. Refer to Chapter 5.

Note:

If a connection to

Mains Earth

is provided by means other than through the

line cord earth conductor (such as through a serial port or parallel port ground
connection), the 601PRO will not detect an Open Earth Error as part of the
outlet test that occurs upon power-up.