Errors and suggested corrective actions – Fluke Biomedical 601PRO XL User Manual
Page 162

6 0 1 P R O S E R I E S
X L
1 1 - 2
2. Errors and Suggested Corrective Actions
Follow the suggested corrective action for errors in the range of 0A00-1900. If the error
persists, you should collect as much information as possible about what you were doing at the
time of the error, and contact Fluke Biomedical or your local Service Center to report the
problem.
Code Description
Corrective
Action
0A00
System Test: 500-Volt
failure.
Recycle power and try again.
0B00
System Test: Mains
Voltage failure.
Recycle power and try again.
0C00
System Test: No Earth
connection.
Try a different wall outlet (see note below).
0D00
System Test Failure
preventing action.
The unit failed the power-up test(s). Recycle power and try again.
0E00
A to D read failure.
Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.
0F00
Calibration Value out of
range.
Confirm that patient leads are disconnected. If the problem
persists, contact Fluke Biomedical or your local Service Center to
report the problem.
1000
I Scale failure during
calibration.
Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.
1100
Voltage across 0.5 Ohm
cal resistor error.
Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.
1200 Calibration
Resistor
current error.
Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.
1300
System test leakage error.
Internal hardware error. Contact Fluke Biomedical or your local
Service Center to report the problem.
1400
DIR checksum error.
Contact Fluke Biomedical or your local Service Center to report
the problem. The most probable cause is that Flash is worn out in
that sector.
1500
Result checksum error.
Contact Fluke Biomedical or your local Service Center to report
the problem. The most probable cause is that Flash is worn out in
that sector.
1600
No Test. Results storage
full. Delete old.
Delete some stored results to make room for new ones. Refer to
Chapter 5.
1700
Printer timeout. Print
buffer full.
You need to have a printer enabled and connected. Refer to
Chapter 2.
1800
Device Record Format is
invalid.
Review the format of the data being transmitted to the 601PRO.
Refer to Chapter 6.
1900
Serial Port timeout.
Cannot transmit.
Verify that you are using the correct serial cable, that it is properly
connected, and that the communications parameters are correctly
set-up on both the PC and the 601. Refer to Chapter 6.
2100
Cannot delete all Device
information until all test
results have been deleted.
Delete all test results. Refer to Chapter 5.
☛
Note:
If a connection to
Mains Earth
is provided by means other than through the
line cord earth conductor (such as through a serial port or parallel port ground
connection), the 601PRO will not detect an Open Earth Error as part of the
outlet test that occurs upon power-up.