8 - 5

IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests

DC-Only tests will only be performed if DC readings are enabled. The types (B, BF, and CF) represent applied
parts for Insulation Resistance, Patient Leakage Current, Patient Auxiliary, and Mains on Applied Part.

IEC 601-1 STANDARD

DUT OUTLET

CLASS I

CLASS II

CLASS IP

DESCRIPTION POLARIT

Y

CIRCUIT B BF CF B BF CF B BF CF

Line Voltage V

N/A

OFF

NL

NL

NL NL NL NL

NT

NT

NT

Protective Earth
Resistance

**

N/A OFF

0.2

0.2

0.2

NT NT

NT

NT

NT

NT

Insulation Resistance
L1-L2-Case M

***

N/A OFF 2 2 2 NT NT

NT

NT

NT

NT

Insulation Resistance
All Case M

N/A OFF

NT

2 2 NT NT

NT

NT

NT

NT

Startup Delay (Auto
only)

Norm

Norm

Current in L2 -- A

Norm

Norm

NL

NL

NL

NL

NL

NL

NT

NT

NT

Enclosure Leakage --
µA

Norm Norm 100 100 100 100 100 100 NT NT NT

Patient Leakage
Current --

µA

Norm Norm

100

100

10 100 100 10

100

100

10

Patient Leakage
Current --

µA – DC

Only

Norm Norm 10 10 10 10 10 10

10 10 10

Patient Auxiliary --

µA

Norm Norm

100

100

10 100 100 10

100

100

10

Patient Auxiliary --

µA -

DC Only

Norm Norm 10 10 10 10 10 10

10 10 10

Mains on AP –

µA ****

Norm Norm NT 5000 50 NT 5000 50 NT 5000 50

Earth Leakage --

µA

Norm No

E

500

500

500 NT NT

NT

NT

NT

NT

Enclosure Leakage --
µA

Norm No

E

500

500

500 NT NT

NT

NT

NT

NT

Patient Leakage
Current --

µA

Norm No

E

500

500

50 NT NT

NT

NT

NT

NT

Patient Leakage
Current --

µA – DC

Only

Norm No

E 50 50 50 NT NT

NT

NT

NT

NT

Patient Auxiliary --

µA

Norm No

E

500

500

50 NT NT

NT

NT

NT

NT

Patient Auxiliary --

µA -

DC Only

Norm No

E 50 50 50 NT NT

NT

NT

NT

NT

Power Off Delay

Norm

No L2

No E

Earth Leakage --

µA

Norm No

L2

No E

1000 1000 1000 NT

NT NT NT NT NT

Enclosure Leakage --
µA

Norm No

L2 500 500 500 500 500 500 NT NT NT

Patient Leakage
Current --

µA

Norm No

L2

500

500 50 500 500 50

NT

NT

NT

Patient Leakage
Current --

µA – DC

Only

Norm No

L2 50 50 50 50 50 50

NT NT NT

Patient Auxiliary --

µA

Norm No

L2

500

500 50 500 500 50

NT

NT

NT

Patient Auxiliary --

µA -

DC Only

Norm No

L2 50 50 50 50 50 50

NT NT NT

Startup Delay

Reverse

Norm

Enclosure Leakage --
µA

Reverse Norm 100 100 100 100 100 100 NT NT NT

Patient Leakage
Current --

µA

Reverse Norm 100 100 10 100 100 10 NT NT NT

Patient Leakage
Current --

µA – DC

Only

Reverse Norm 10 10 10 10 10 10

NT NT NT

Patient Auxiliary --

µA

Reverse Norm 100 100 10 100 100 10 NT NT NT

Patient Auxiliary --

µA -

DC Only

Reverse Norm 10 10 10 10 10 10

NT NT NT

Mains on AP –

µA *

Reverse Norm NT 5000 50 NT 5000 50 NT NT NT

Earth Leakage --

µA

Reverse No

E 500 500 500 NT NT NT

NT NT NT

Enclosure Leakage --
µA

Reverse No

E 500 500 500 NT NT NT

NT NT NT

Patient Leakage
Current --

µA

Reverse No

E 500 500 50 NT NT NT

NT NT NT