Fluke Biomedical 601PRO XL User Manual
Page 119

S T A N D A R D S A N D P R I N C I P L E S
8 - 5
IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests
DC-Only tests will only be performed if DC readings are enabled. The types (B, BF, and CF) represent applied
parts for Insulation Resistance, Patient Leakage Current, Patient Auxiliary, and Mains on Applied Part.
IEC 601-1 STANDARD
DUT OUTLET
CLASS I
CLASS II
CLASS IP
DESCRIPTION POLARIT
Y
CIRCUIT B BF CF B BF CF B BF CF
Line Voltage V
N/A
OFF
NL
NL
NL NL NL NL
NT
NT
NT
Protective Earth
Resistance
Ω
**
N/A OFF
0.2
0.2
0.2
NT NT
NT
NT
NT
NT
Insulation Resistance
L1-L2-Case M
Ω
***
N/A OFF 2 2 2 NT NT
NT
NT
NT
NT
Insulation Resistance
All Case M
Ω
N/A OFF
NT
2 2 NT NT
NT
NT
NT
NT
Startup Delay (Auto
only)
Norm
Norm
Current in L2 -- A
Norm
Norm
NL
NL
NL
NL
NL
NL
NT
NT
NT
Enclosure Leakage --
µA
Norm Norm 100 100 100 100 100 100 NT NT NT
Patient Leakage
Current --
µA
Norm Norm
100
100
10 100 100 10
100
100
10
Patient Leakage
Current --
µA – DC
Only
Norm Norm 10 10 10 10 10 10
10 10 10
Patient Auxiliary --
µA
Norm Norm
100
100
10 100 100 10
100
100
10
Patient Auxiliary --
µA -
DC Only
Norm Norm 10 10 10 10 10 10
10 10 10
Mains on AP –
µA ****
Norm Norm NT 5000 50 NT 5000 50 NT 5000 50
Earth Leakage --
µA
Norm No
E
500
500
500 NT NT
NT
NT
NT
NT
Enclosure Leakage --
µA
Norm No
E
500
500
500 NT NT
NT
NT
NT
NT
Patient Leakage
Current --
µA
Norm No
E
500
500
50 NT NT
NT
NT
NT
NT
Patient Leakage
Current --
µA – DC
Only
Norm No
E 50 50 50 NT NT
NT
NT
NT
NT
Patient Auxiliary --
µA
Norm No
E
500
500
50 NT NT
NT
NT
NT
NT
Patient Auxiliary --
µA -
DC Only
Norm No
E 50 50 50 NT NT
NT
NT
NT
NT
Power Off Delay
Norm
No L2
No E
Earth Leakage --
µA
Norm No
L2
No E
1000 1000 1000 NT
NT NT NT NT NT
Enclosure Leakage --
µA
Norm No
L2 500 500 500 500 500 500 NT NT NT
Patient Leakage
Current --
µA
Norm No
L2
500
500 50 500 500 50
NT
NT
NT
Patient Leakage
Current --
µA – DC
Only
Norm No
L2 50 50 50 50 50 50
NT NT NT
Patient Auxiliary --
µA
Norm No
L2
500
500 50 500 500 50
NT
NT
NT
Patient Auxiliary --
µA -
DC Only
Norm No
L2 50 50 50 50 50 50
NT NT NT
Startup Delay
Reverse
Norm
Enclosure Leakage --
µA
Reverse Norm 100 100 100 100 100 100 NT NT NT
Patient Leakage
Current --
µA
Reverse Norm 100 100 10 100 100 10 NT NT NT
Patient Leakage
Current --
µA – DC
Only
Reverse Norm 10 10 10 10 10 10
NT NT NT
Patient Auxiliary --
µA
Reverse Norm 100 100 10 100 100 10 NT NT NT
Patient Auxiliary --
µA -
DC Only
Reverse Norm 10 10 10 10 10 10
NT NT NT
Mains on AP –
µA *
Reverse Norm NT 5000 50 NT 5000 50 NT NT NT
Earth Leakage --
µA
Reverse No
E 500 500 500 NT NT NT
NT NT NT
Enclosure Leakage --
µA
Reverse No
E 500 500 500 NT NT NT
NT NT NT
Patient Leakage
Current --
µA
Reverse No
E 500 500 50 NT NT NT
NT NT NT