Shortcut key 1: current consumption test – Fluke Biomedical 601PRO XL User Manual
Page 57

M A N U A L M O D E
3 - 1 7
Modify the configuration of the front panel outlet by pressing the appropriate
SOFT
KEY
on the 601PRO:
•
SOFT KEY 1:
Toggles the polarity from Normal to Off to Reverse Polarity.
•
SOFT KEY 2
: Changes the earth line from Earth (closed earth) to No Earth (open
earth).
•
SOFT KEY 3
: Changes the L2 line from L2 (closed L2) to No L2 (open L2).
•
SOFT KEY 4:
Returns to the
Mains Voltage
menu.
Shortcut Key 1: Current Consumption Test
To measure the current (in amperes) flowing in L2 of the DUT, press
shortcut key 1
,
or select the Current Consumption Test by pressing
TESTS/AUTOMODES
, then
MANUAL
, then
*MORE
until the test is displayed.
To measure the current consumption:
• From the
MAIN MENU,
or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Press
shortcut key 1
. The
Current Consumption
test is displayed, and the test
begins immediately. Display values are continuously updated until another test is
selected.
baw049f.eps
• Modify the configuration of the front panel outlet by pressing the appropriate
SOFT KEY
on the 601PRO.
•
SOFT KEY 1
toggles the DUT outlet Polarity from Normal to Off to Reverse.
•
SOFT KEY 2
toggles the DUT outlet from Earth to No Earth.
•
SOFT KEY 3
toggles the DUT outlet from L2 to No L2.
• Press the
print data
key at any time to generate a printout of the latest
measurement(s)
.