M A N U A L M O D E

3 - 1 7

Modify the configuration of the front panel outlet by pressing the appropriate

SOFT

KEY

on the 601PRO:

SOFT KEY 1:

Toggles the polarity from Normal to Off to Reverse Polarity.

SOFT KEY 2

: Changes the earth line from Earth (closed earth) to No Earth (open

earth).

SOFT KEY 3

: Changes the L2 line from L2 (closed L2) to No L2 (open L2).

SOFT KEY 4:

Returns to the

Mains Voltage

menu.

Shortcut Key 1: Current Consumption Test

To measure the current (in amperes) flowing in L2 of the DUT, press

shortcut key 1

,

or select the Current Consumption Test by pressing

TESTS/AUTOMODES

, then

MANUAL

, then

*MORE

until the test is displayed.

To measure the current consumption:

• From the

MAIN MENU,

or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Press

shortcut key 1

. The

Current Consumption

test is displayed, and the test

begins immediately. Display values are continuously updated until another test is

selected.

baw049f.eps

• Modify the configuration of the front panel outlet by pressing the appropriate

SOFT KEY

on the 601PRO.

SOFT KEY 1

toggles the DUT outlet Polarity from Normal to Off to Reverse.

SOFT KEY 2

toggles the DUT outlet from Earth to No Earth.

SOFT KEY 3

toggles the DUT outlet from L2 to No L2.

• Press the

print data

key at any time to generate a printout of the latest

measurement(s)

.