M A N U A L M O D E
3 - 1 9
To perform an
Insulation Resistance Applied Part-to-Case
test, press
AP INSUL
while the
Insulation Resistance
test is displayed. In this test, 500 VDC is applied
between (one or all in common) applied parts and the DUT Earth.
• Attach the device's applied parts to the 601PRO's applied part terminals.
• Press
APPLIED PART (SOFT KEY 4)
to select the applied part(s) to be tested.
• Press
START TEST
to begin the test.
baw051f.eps
⇒
SOFT KEY 1
starts or stops the test.
⇒
SOFT KEY 2
changes the test to Insulation Resistance.
⇒
SOFT KEY 4
toggles the applied part from:
• ALL to Case
• RA to Case
• RL to Case
• LA to Case
• LL to Case
• V1-V6 to Case
⇒ The unit will beep repeatedly as long as the test is active.
⇒ Press
STOP TEST (SOFT KEY 1)
to terminate the test.
• Press the
print data
key at any time to generate a printout of the latest
measurement(s)
.
Important:
If all of the applied parts correspond to the instrument type, the applied
parts will be tied together and one reading will be taken. If any of the
applied parts differ from the instrument type, all applied parts will be
tested individually, based on the type of applied part. This applies to
Auto
and
Step
modes only.