beautypg.com

Fluke Biomedical 601PRO XL User Manual

Page 122

background image

6 0 1 P R O S E R I E S

X L

8 - 8

IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests (Continued)

IEC 601-1 STANDARD

DUT OUTLET

CLASS I

CLASS II

CLASS IP

DESCRIPTION

POLARITY

CIRCUIT B BF CF B BF CF B BF CF

Startup Delay

Reverse

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Leakage Current -
µA

Reverse No

E 500 500 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Norm Norm 10 10 10 10 10 10 10 10 10

Power Off Delay

Norm

No L2

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Norm No

L2

50

50

50

50

50

50

NT

NT

NT

Startup Delay

Norm

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Norm

No

E 50 50 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Reverse Norm 10 10 10 10 10 10 NT NT NT

Power Off Delay

Reverse

No L2

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Reverse No

L2 50 50 50 50 50 50 NT NT NT

Startup Delay

Reverse

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Leakage Current -
-

µA -DC Only

Reverse No

E 50 50 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Mains on AP --

µA ****

Norm

Norm NT 5000 50 NT 5000 50 5000

5000 50

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Mains on AP --

µA *

Reverse Norm NT

5000

50 NT

5000

50 NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Patient Auxiliary --

µA

Norm

Norm 100 100 10 100 100 10 100 100 10

Power Off Delay

Norm

No L2

N/A

N/A

N/A

Patient Auxiliary --

µA

Norm No

L2

500

500

50

500

500

50

NT

NT

NT

Startup Delay

Norm

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Auxiliary --

µA

Norm

No

E 500 500 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Patient Auxiliary --

µA

Reverse Norm 100

100

10 100

100 10 NT NT NT

Power Off Delay

Reverse

No L2

N/A

N/A

N/A

Patient Auxiliary --

µA

Reverse No

L2 500

500 50 500

500 50 NT NT NT

Startup Delay

Reverse

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Auxiliary --

µA

Reverse No

E 500 500 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Patient Auxiliary --

µA DC

Only

Norm Norm 10 10 10 10 10 10 10 10 10

Power Off Delay

Norm

No L2

N/A

N/A

N/A

Patient Auxiliary --

µA DC

Only

Norm No

L2

50

50

50

50

50

50

NT

NT

NT

Startup Delay

Norm

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Auxiliary --

µA DC

Only

Norm

No

E 50 50 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Patient Auxiliary --

µA DC

Only

Reverse Norm 10 10 10 10 10 10 NT NT NT