Computer control commands (cont.) – Fluke Biomedical 601PRO XL User Manual
Page 157

C O M P U T E R C O N T R O L
1 0 - 9
Computer Control Commands (Cont.)
Patient Leakage Measurement – DC-Only Commands (OUTLET AUTOMATICALLY ON FOR TEST)
COMMAND DESCRIPTION
NOTES
[PALD]
Pat. Leakage ALL-EARTH
601 Test Load
[PRAD]
Pat. Leakage RA-EARTH
601 Test Load
[PRLD]
Pat. Leakage RL-EARTH
601 Test Load
[PLAD]
Pat. Leakage LA-EARTH
601 Test Load
[PLLD]
Pat. Leakage LL-EARTH
601 Test Load
[PV1D]
Pat. Leakage V1-V6-EARTH
601 Test Load
Auxiliary Current Measurement – DC-Only Commands (OUTLET AUTOMATICALLY ON FOR TEST)
COMMAND DESCRIPTION
NOTES
[ARAD]
Aux. Leakage RA-ALL
601 Test Load
[ARLD]
Aux. Leakage RL-ALL
601 Test Load
[ALAD]
Aux. Leakage LA-ALL
601 Test Load
[ALLD]
Aux. Leakage LL-ALL
601 Test Load
[AV1D]
Aux. Leakage V1-V6-ALL
601 Test Load
IEC 1010 Tests (OUTLET AUTOMATICALLY ON FOR TEST)
COMMAND DESCRIPTION
NOTES
[ACV] Accessible
Voltage
[ACL]
Accessible Leakage RMS
IEC 1010 Test Load
[ACLD]
Accessible Leakage DC
IEC 1010 Test Load
VDE Tests
COMMAND DESCRIPTION
NOTES
[EDL]
Equivalent Device Leak
601 Test Load
[EPL]
Equivalent Patient Leakage RMS
601 Test Load
[EPLD]
Equivalent Patient Leakage DC
601 Test Load