S T A N D A R D S A N D P R I N C I P L E S

8 - 1 5

Test Principle: Protective Earth Resistance

Notes:

S1 = Not applicable
S2 = Not applicable

S3 = Open
S5 = Open

baw137f.eps

Function:

Measures the impedance between the 601PRO’s outlet PE terminal and exposed conductive parts of

the DUT that are connected to the DUT’s protective earth. Test current is applied for approximately 5

seconds.

Applicable to:

IEC 601-1, Class I:

Current source = 25 A ± 10%

VDE 751-1, Class I:

Current source = 5-25 A

HEI 95, Class I:

Current source = 1 A

VDE 701, Class I:

Current source = 1 A

IEC 1010, Class I:

Current source = 1 A

AAMI:

Current source = 1 A

AS/NZS 3551:

Current source = 1 A