2. Tests Available in a Custom Standard
The following table lists all of the tests available in a custom standard, in the order in
which the operator will be prompted. It also indicates whether the test requires limits.
Custom Test Standard Prompt Order
SCREEN #
TEST
TEST LIMIT RANGE
1
Mains Voltage
NO LIMIT
2
Earth Resistance
0-3000 mOhm
3
Insul. Res: L1-L2-CASE
0-400 Mohm
4
Insul. Res: AP-CASE
0-400 Mohm
5
Current Consumption
NO LIMIT
6
Earth Leakage: Normal
0-8000 µA
7
Earth Leakage: No L2
0-8000 µA
8
Earth Leakage: Rev
0-8000 µA
9
Earth Leakage: Rev, No L2
0-8000 µA
10
Enclosure Leakage: Normal
0-8000 µA
11
Enclosure Leakage DC: Normal
0-8000 µA
12
Enclosure Leakage: No L2
0-8000 µA
13
Enclosure Leakage DC: No L2
0-8000 µA
14
Enclosure Leakage: No Earth
0-8000 µA
15
Enclosure Leakage DC: No Earth
0-8000 µA
16
Enclosure Leakage: Rev
0-8000 µA
17
Enclosure Leakage DC: Rev
0-8000 µA
18
Enclosure Leakage: Rev, No L2
0-8000 µA
19
Enclosure Leakage DC: Rev, No L2
0-8000 µA
20
Enclosure Leakage: Rev, No Earth
0-8000 µA
21
Enclosure Leakage DC: Rev, No Earth
0-8000 µA
22
Equiv Device Leakage
0-8000 µA
23
Patient Leakage: Normal
0-8000 µA
24
Patient Leakage DC: Normal
0-8000 µA
25
Patient Leakage: No L2
0-8000 µA
26
Patient Leakage DC: No L2
0-8000 µA
27
Patient Leakage: No Earth
0-8000 µA
28
Patient Leakage DC: No Earth
0-8000 µA
29
Patient Leakage: Rev
0-8000 µA
30
Patient Leakage DC: Rev
0-8000 µA
31
Patient Leakage: Rev, No L2
0-8000 µA
32
Patient Leakage DC: Rev, No L2
0-8000 µA
33
Patient Leakage: Rev, No Earth
0-8000 µA
34
Patient Leakage DC: Rev, No Earth
0-8000 µA
35
Mains On Applied Part: Normal
0-8000 µA
36
Mains On Applied Part: Rev
0-8000 µA
37
Patient Aux Current: Normal
0-8000 µA
38
Patient Aux Current DC: Normal
0-8000 µA
39
Patient Aux Current: No L2
0-8000 µA
40
Patient Aux Current DC: No L2
0-8000 µA
41
Patient Aux Current: No Earth
0-8000 µA
42
Patient Aux Current DC: No Earth
0-8000 µA
43
Patient Aux Current: Rev
0-8000 µA
44
Patient Aux Current DC: Rev
0-8000 µA
45
Patient Aux Current: Rev, No L2
0-8000 µA
46
Patient Aux Current DC: Rev, No L2
0-8000 µA
47
Patient Aux Current: Rev, No Earth
0-8000 µA
48
Patient Aux Current DC: Rev, No Earth
0-8000 µA
49
Equiv Patient Leakage
0-8000 µA
50
Accessible Voltage: Normal
0-300 V
51
Accessible Voltage: No Earth
0-300 V
52
Accessible Voltage: No L2
0-300 V
53
Accessible Voltage: Rev
0-300 V
54
Accessible Voltage: Rev, No Earth
0-300 V
55
Accessible Voltage: Rev, No L2
0-300 V