Tests available in a custom standard – Fluke Biomedical 601PRO XL User Manual

Page 144

background image

6 0 1 P R O S E R I E S

X L

9 - 4

2. Tests Available in a Custom Standard

The following table lists all of the tests available in a custom standard, in the order in
which the operator will be prompted. It also indicates whether the test requires limits.


Custom Test Standard Prompt Order

SCREEN #

TEST

TEST LIMIT RANGE

1

Mains Voltage

NO LIMIT

2

Earth Resistance

0-3000 mOhm

3

Insul. Res: L1-L2-CASE

0-400 Mohm

4

Insul. Res: AP-CASE

0-400 Mohm

5

Current Consumption

NO LIMIT

6

Earth Leakage: Normal

0-8000 µA

7

Earth Leakage: No L2

0-8000 µA

8

Earth Leakage: Rev

0-8000 µA

9

Earth Leakage: Rev, No L2

0-8000 µA

10

Enclosure Leakage: Normal

0-8000 µA

11

Enclosure Leakage DC: Normal

0-8000 µA

12

Enclosure Leakage: No L2

0-8000 µA

13

Enclosure Leakage DC: No L2

0-8000 µA

14

Enclosure Leakage: No Earth

0-8000 µA

15

Enclosure Leakage DC: No Earth

0-8000 µA

16

Enclosure Leakage: Rev

0-8000 µA

17

Enclosure Leakage DC: Rev

0-8000 µA

18

Enclosure Leakage: Rev, No L2

0-8000 µA

19

Enclosure Leakage DC: Rev, No L2

0-8000 µA

20

Enclosure Leakage: Rev, No Earth

0-8000 µA

21

Enclosure Leakage DC: Rev, No Earth

0-8000 µA

22

Equiv Device Leakage

0-8000 µA

23

Patient Leakage: Normal

0-8000 µA

24

Patient Leakage DC: Normal

0-8000 µA

25

Patient Leakage: No L2

0-8000 µA

26

Patient Leakage DC: No L2

0-8000 µA

27

Patient Leakage: No Earth

0-8000 µA

28

Patient Leakage DC: No Earth

0-8000 µA

29

Patient Leakage: Rev

0-8000 µA

30

Patient Leakage DC: Rev

0-8000 µA

31

Patient Leakage: Rev, No L2

0-8000 µA

32

Patient Leakage DC: Rev, No L2

0-8000 µA

33

Patient Leakage: Rev, No Earth

0-8000 µA

34

Patient Leakage DC: Rev, No Earth

0-8000 µA

35

Mains On Applied Part: Normal

0-8000 µA

36

Mains On Applied Part: Rev

0-8000 µA

37

Patient Aux Current: Normal

0-8000 µA

38

Patient Aux Current DC: Normal

0-8000 µA

39

Patient Aux Current: No L2

0-8000 µA

40

Patient Aux Current DC: No L2

0-8000 µA

41

Patient Aux Current: No Earth

0-8000 µA

42

Patient Aux Current DC: No Earth

0-8000 µA

43

Patient Aux Current: Rev

0-8000 µA

44

Patient Aux Current DC: Rev

0-8000 µA

45

Patient Aux Current: Rev, No L2

0-8000 µA

46

Patient Aux Current DC: Rev, No L2

0-8000 µA

47

Patient Aux Current: Rev, No Earth

0-8000 µA

48

Patient Aux Current DC: Rev, No Earth

0-8000 µA

49

Equiv Patient Leakage

0-8000 µA

50

Accessible Voltage: Normal

0-300 V

51

Accessible Voltage: No Earth

0-300 V

52

Accessible Voltage: No L2

0-300 V

53

Accessible Voltage: Rev

0-300 V

54

Accessible Voltage: Rev, No Earth

0-300 V

55

Accessible Voltage: Rev, No L2

0-300 V