Fluke Biomedical 601PRO XL User Manual

Page 50

background image

6 0 1 P R O S E R I E S

X L

3 - 1 0

CONDITION 3:
If at least one Applied Part is defined with a DIFFERENT Type than any of

the leads or the EQUIPMENT Type, then the Applied Parts will be tested

INDIVIDUALLY.

1. To view or change the lead types, press the

view present settings

key

while at the

MAIN MENU

.

2. Press

EDIT LEADS

at the following menu:

baw037f.eps

3. Enter the

Number of Applied Part Leads

using the numeric keys:

baw038f.eps


This selection allows the user to define the number of applied parts to be
tested during the Mains on Applied Part, Insulation Resistance, Patient
Auxiliary Current, and Patient Leakage tests. This screen will only
appear when the selected test standard is one of the following:

♦ IEC 601-1
♦ HEI 95
♦ Custom 1-4*
♦ AAMI*
♦ AS/NZS 3551

Press

enter

.

*Note:

When using the AAMI or Custom test standard, only the applied

parts can be assigned; the Type cannot be changed.