Saving standard, class, type and test current – Fluke Biomedical 601PRO XL User Manual
Page 46

6 0 1 P R O S E R I E S
X L
3 - 6
• At the 601PRO
MAIN MENU
, press
CLASS/TYPE
:
baw005f.eps
Press the
SOFT KEY
below the desired
CLASS/TYPE
, or press
*MORE
for additional
options.
baw006f.eps
☛
Note: When
the
CLASS/TYPE
is changed, the lead types are not changed.
☛
Note:
Once the selection is made, the 601PRO will automatically return to the
MAIN MENU
.
5. Saving Standard, Class, Type and Test Current
The Standard, Class, and Type can be saved in non-volatile memory by running a
manual test that is applicable to that configuration. For example, make a selection
under
VIEW PRESENT SETTINGS
menu. Next, press the
0
button to run a mains
Voltage test. Press
PREVIOUS
to end the test and now the parameter will be saved. It
will remain as the default after cycling power off and back on.
☛
Note:
With units operating with V2.00 or later firmware, the Protective Earth
Resistance Test Current will also be saved as above by running a manual
test.