beautypg.com

Replacing the paper roll, Printer self-test, Seiko thermal printer (beige chassis) – Fluke Biomedical 601PRO XL User Manual

Page 178

background image

6 0 1 P R O S E R I E S

X L

C - 2

Paper will automatically
feed when inserted.

PRINTER HOUSING

(inside unit chassis)

PAPER PORT

ROLL CAP

PAPER ROLL

baw157f.eps

Replacing the paper roll

1. Remove the roll cap by squeezing the barbs at the end of the paper roll

shaft.

2. Load a new roll (PN 2248743) oriented as shown in the figure above.

3. Feed a clean, straight edge from the roll into the entry port of the printer

mechanism. The paper will be sensed automatically and feed into the front
slot for printing.

4. Wind any excess paper onto the roll and replace the roll cap.

5. Close the printer cover.

Printer Self-Test

1. Turn off the power to the 601Pro analyzer.

2. Press and hold the paper feed button on the printer cover.

3. Turn the power back on.

4. Release the paper feed button.

Seiko Thermal Printer (beige chassis)
Before proceeding, turn off the power to the 601PRO analyzer.

To gain access to the printer, gently press the latch buttons on the sides of the front
panel and remove the panel.