beautypg.com

Shortcut key 8: patient auxiliary current test – Fluke Biomedical 601PRO XL User Manual

Page 68

background image

6 0 1 P R O S E R I E S

X L

3 - 2 8

Important:

If all of the applied parts correspond to the instrument type, the applied

parts will be tied together and one reading will be taken. If any of the

applied parts differ from the instrument type, all applied parts will be

tested individually, based on the type of applied part. This applies to

Auto

and

Step

modes only.

Shortcut Key 8: Patient Auxiliary Current Test

Patient Auxiliary currents are measured between any selected ECG jack and the

remaining selected ECG jacks (refer to Lead Type Definitions on page 3-9). All

measurements may have either a true RMS or a DC-only response.

The following outlet conditions apply when performing this test:

⇒ Normal Polarity, Earth Open, Outlet ON

⇒ Normal Polarity, Outlet ON

⇒ Normal Polarity, L2 Open, Outlet ON

⇒ Reversed Polarity, Outlet ON

⇒ Reversed Polarity, Earth Open, Outlet ON

⇒ Reversed Polarity, L2 Open, Outlet ON

Note:

If the current test standard is IEC 1010, then the Patient Auxiliary

Current test will not be available. Refer to Chapter 8, Standards and

Principles.

To perform a

Patient Auxiliary Current

test:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach the patient leads to the 601PRO ECG jacks.
• Define the Lead Types from the View Settings Option (refer to: Lead Type

Definitions in Section 5 of this chapter).

• Press

shortcut key 8

. The

Patient Auxiliary Current

test is displayed, and the

test begins immediately. Display values are continuously updated until another

test is selected.

baw060f.eps