Fluke Biomedical 601PRO XL User Manual

Page 67

background image

M A N U A L M O D E

3 - 2 7

• Select the desired outlet configuration and applied part to test using the

appropriate

SOFT KEYS

:

SOFT KEY 1

takes one pair of readings and displays the result.

SOFT KEY 2

allows the user to perform a calibration.

SOFT KEY 3

toggles the DUT outlet Polarity from Normal to DUT Off to

Reverse Polarity, or Reverse to DUT Off to Normal Polarity.

SOFT KEY 4

toggles the applied part from:

• ALL to Earth

• RA to Earth

• RL to Earth

• LA to Earth

• LL to Earth

• V1-V6 to Earth

• Press

START TEST (SOFT KEY 1)

to begin the test.

WARNING: High voltage is present at applied part terminals while measurements are

being taken.

• Press the

print data

key to generate a printout of the latest measurement.

To perform a calibration from the

Mains on Applied Part

test, press CAL

(SOFT KEY 2)

.

• Disconnect ALL patient leads, test leads, and DUT outlet connections.

• Press

CAL

to begin calibration, as shown:

baw059f.eps

WARNING: A 2-beep-per-second signal indicates high voltage present at the applied part
terminals while a calibration is being performed.

• When the calibration is finished, the

Mains on Applied Part

test will reappear.

Note:

If the calibration fails, the previously stored readings will be used until a

passing calibration has occurred. Also, the

esc/stop

key has no effect

during calibration.