S T A N D A R D S A N D P R I N C I P L E S
8 - 2 5
Test Principle: Accessible Voltage
Notes:
S1 = Variable
S2 = Variable
S3 = Variable
S5 = Closed
baw147f.eps
Function:
Measures the voltage present on accessible parts relative to mains PE. The test is repeated with
normal and reversed mains polarity and single faults of no earth and no L2.
Applicable to:
IEC 1010