M A N U A L M O D E

3 - 2 5

To perform a

Patient Leakage Current

test using the

test-shortcut

keys:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach the applied parts to the 601PRO's applied part terminals.

Press

shortcut key 6

. The

Patient Leakage

test is displayed, and the test begins

immediately.

baw057f.eps

SOFT KEY 1

toggles the DUT outlet Polarity from Normal to DUT Off to

Reverse Polarity, or Reverse to DUT Off to Normal Polarity.

SOFT KEY 2

toggles the DUT outlet from Earth to No Earth.

SOFT KEY 3

toggles the DUT outlet from L2 to No L2.

SOFT KEY 4

toggles the applied part from:

ALL to Earth (RMS)

RA to Earth (RMS)

RL to Earth (RMS)

LA to Earth (RMS)

LL to Earth (RMS)

V1-V6 to Earth (RMS)

ALL to Earth DC

RA to Earth DC

RL to Earth DC

LA to Earth DC

LL to Earth DC

V1-V6 to Earth DC


Note:

If the current test standard being used does not include Patient Leakage

DC readings, or the DC option is not enabled, then DC readings will not

be available through the

APPLIED PART SOFT KEY

selections. Refer to

Chapter 8, Standards and Principles.