Test Data Record ASCII Character Formats

Test data records must only contain ASCII characters formatted as shown in the table

below. The first portion of the record contains information about the test setup, time, and

date, etc., and each of these fields is surrounded by quotation marks. The remainder of the

information pertains to test results: All lines are delimited with . The data

fields are delimited by commas as shown. The "P/F/N" relates to the Pass/Failure or test

NOT performed. The individual safety tests are represented by numbers in the first field of

each line and are described in parenthesis. Fields left blank at the time of test will appear

as (a blank line)

FIELD DESCRIPTION

601PRO Serial Number

15 characters max

Month of test

“XX” two digit month

Day of test

“XX” two digit day

Year of test

“XX” two digit year

Hour of test

“XX” two digit (24 hour clock)

Minute of test

“XX” two digit

Technician

15 characters max

Control Number

15 characters max

Procedure ID

20 characters max

Serial Number of DUT

16 characters max

Location

30 characters max

Device Type

20 characters max

Manufacturer

25 characters max

Class of test

“I”, “II”, “IP”, “??”

Type of test

“B”, “BF”, “CF”, “T”, “L”, “??”

Test Standard

“IEC601-1”, “IEC1010”, “VDE751-1”, “VDE701”, “AAMI”,
“AS/NZS 3551”, “CUSTOM1”, “CUSTOM2”, “CUSTOM3”
and “CUSTOM4”

Test Current

“1”, "10", or “25” Amperes

Leads

“X” ( 0,1,2,3, 4 or 5 )

Lead Definition

“(RA-XX)(RL-XX)(LA-XX)(LL-XX)(V-XX)”

Power Off Delay

“0-9999”

Startup Delay

“0-9999”

”100”,”data”

(Voltage L1-Earth)

”101”,”data”

(Voltage L2-Earth)

”102”,”data”

(Voltage L1-L2)

”150”,”data”,”P/F/N” (Earth

Resistance)

”200”,”data”,”P/F”

(Insulation Resistance)

”201”,”data”,”P/F”

(Insulation Resistance AP- All leads )

“205” ,”data”,”P/F”

(Insulation Resistance AP-RA only)

“206” ,”data”,”P/F”

(Insulation Resistance AP-RL only)

“207” ,”data”,”P/F”

(Insulation Resistance AP-LA only)

“208” ,”data”,”P/F”

(Insulation Resistance AP-LL only)

“209” ,”data”,”P/F”

(Insulation Resistance AP-V1-V6 only)

”250”,”data”

(Current Consumption)

”300”,”data”,”P/F/N”

(Earth Leakage Normal)

”301”,”data”,”P/F/N”

(Earth Leakage No L2)

”302”,”data”,”P/F/N”

(Earth Leakage Rev Pol)

”303”,”data”,”P/F/N”

(Earth Leakage Rev No L2)

”350”,”data”,”P/F/N”

(Enclosure Leak Normal)