Test principle: mains on applied part – Fluke Biomedical 601PRO XL User Manual
Page 134

6 0 1 P R O S E R I E S
X L
8 - 2 0
Test Principle: Mains on Applied Part
Notes:
S1 = Closed
S2 = Variable
S3 = Closed
S4 = Variable
S5 = Closed
S6 = Variable
baw142f.eps
Function:
Measures the current that flows in response to an isolated AC voltage applied between selected
applied parts and earth (and any conductive part connected to the RED terminal). The magnitude of
the voltage is at least 110% of the mains voltage. The test is performed with normal and reverse
mains supply polarity using S2, and normal and reverse isolated voltage supply polarity using S4.
Applicable to:
IEC 601-1
AAMI
AS/NZS
3551
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)