6 0 1 P R O S E R I E S

X L

8 - 6

IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests (Continued)

IEC 601-1 STANDARD

DUT OUTLET

CLASS I

CLASS II

CLASS IP

DESCRIPTION POLARITY

CIRCUIT

B

BF

CF B BF

CF B BF

CF

Patient Leakage
Current --

µA – DC

Only

Reverse No

E 50 50 50 NT NT NT NT NT NT

Patient Auxiliary --

µA

Reverse No

E

500

500

50 NT NT NT NT NT NT

Patient Auxiliary --

µA -

DC Only

Reverse No

E 50 50 50 NT NT NT NT NT NT

Power Off Delay

Reverse

No L2

No E

Earth Leakage --

µA

Reverse No

L2

No E

100

0

1000 1000 NT NT

NT NT NT NT

Enclosure Leakage --
µA

Reverse No

L2 500 500 500 500 500 500 NT NT NT

Patient Leakage
Current --

µA

Reverse No

L2

500

500 50 500 500 50 NT NT NT

Patient Leakage
Current --

µA – DC

Only

Reverse No

L2 50 50 50 50 50 50 NT NT NT

Patient Auxiliary --

µA

Reverse No

L2

500

500 50 500 500 50 NT NT NT

Patient Auxiliary –

µA -

DC Only

Reverse No

L2 50 50 50 50 50 50 NT NT NT

*This test takes one minute in Step mode.

**The limit for Protective Earth Resistance is dependent upon the device under test: 0.2

for devices with a non-detachable power supply

cord; 0.1

for devices with an appliance inlet.

***Limits for Insulation Resistance are not specified in IEC 601-1.

****Patient Leakage Current, Mains on Applied Part -- Normal Outlet is performed with NORMAL OUTLET polarity and both Normal and
Reversed phase isolated 110% of mains. Patient Leakage Current, Mains on Applied Part -- Reversed Outlet are performed in the same
manner as with a normal outlet except with REVERSE OUTLET polarity.

The limits have been removed for the Enclosure Leakage test with Class IP due to the fact that zero readings are always expected.

Delays are not applied to Class IP sequences.