beautypg.com

Fluke Biomedical 601PRO XL User Manual

Page 74

background image

6 0 1 P R O S E R I E S

X L

3 - 3 4

WARNING:

A 2-beep-per-second signal indicates high voltage present at the applied

part terminals while a calibration is being performed.

• When the calibration is finished, the

Equivalent Device Leakage

test will

reappear.

Note:

If the calibration fails the previously stored readings will be used until a

passing calibration has occurred. Also, the

esc/stop

key has no effect

during calibration.

Shortcut Key

-

: VDE Equivalent Patient Leakage Test

The Equivalent Patient Leakage Test measures the current flowing from ALL of the

Applied Parts terminals to the shorted L1, L2 and DUT PE in response to an applied

isolated AC voltage greater than 110% of the Mains voltage. Equivalent Device

Leakage test cannot be selected from the manual menu unless the VDE 751-1 or

VDE 701 test standard is selected.

Note:

This test should be performed with a mains voltage > 180 VAC as

specified by the VDE 701 and VDE 751 test standards.

It is recommended that whenever the mains voltage level is changed

(120 VAC to 220 VAC, or 220 VAC to 120 VAC), or when using the

unit for the first time, that a leakage calibration be performed. This can

be done from either of the VDE Equivalent leakage tests or from Mains

on Applied Part.

This test is always performed with all 5 leads.

Important:

The Equivalent Patient Leakage test uses the IEC 601-1 Test Load

exclusively. This is true no matter what the currently defined test

standard is.

The following conditions apply:

⇒ Mains supply to DUT L1 and L2 Open

⇒ DUT L1 and L2 shorted to DUT PE

⇒ 110% of mains voltage is applied between individually selected ECG jacks

connected to applied parts and the 601PRO Earth.