8-1

Standards and Principles

The 601PRO is programmed to use IEC 601-1 as the default test standard. Alternative test

standards

− VDE 751, VDE 701, HEI 95, IEC 1010, AAMI, AS/NZS 3551 and custom − are

available for selection and can be accessed through

SYSTEM SETUP

from the

MAIN MENU

.

Some electrical tests may not be applicable under certain class/type categories (refer to the tables in

this chapter). In these cases, the test will be performed but the test limit will display

Inv

(invalid).

To access the test standards, follow the steps outlined below.

1. Accessing System Setup
2. Selecting the Test Standard
3. Referring to Test Limits for the Selected Standard

Standards and principles for each test may be found on the following pages:

VDE 701 and VDE 751-1

8-4

IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests

8-5

IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests

8-7

AAMI Controlled Power Sequence and Limits

8-10

IEC 1010 Controlled Power Sequence and Limits (IEC 61010-1, Second Ed.)

8-11

HEI 95 Controlled Power Sequence and Test Limits for Auto and Step Tests

8-12

HEI 95 Conventional Test Sequence and Test Limits for Auto and Step Tests

8-13

AS/NZS 3551 Test Sequence and Limits for Auto and Step Tests

8-14

Test Principle diagrams and notes may be found on the following pages:

Protective Earth Resistance

8-15

Earth Leakage Current

8-16

Enclosure Leakage Current

8-17

Mains Insulation Resistance

8-18

Applied Part Insulation Resistance

8-19

Mains on Applied Part

8-20

Patient Auxiliary Current

8-21

Patient Leakage Current

8-22

VDE, Equivalent Device Leakage

8-23

VDE, Equivalent Patient Leakage

8-24

Accessible Voltage

8-25

Accessible Leakage Current

8-26

Chapter

8