Shortcut Key 0: Mains Voltage Test and Dual Lead Voltage Test

This test can be initiated either from the

test-shortcut

keys or the

MAIN MENU

. The

Mains Voltage Test measures and displays the Mains voltages from L1 to Earth, L2

to Earth, and L1 to L2. The Dual Lead Voltage Test measures the voltage between

the RED and BLACK inputs.

Important:

Power to the 601PRO front panel outlet is off during the Mains Voltage

Test, and on during the Dual Lead Voltage Test.

To measure the mains voltage (L1-EARTH, L2-EARTH, L1-L2), press

shortcut key

0

or press

TESTS/AUTOMODES

from the

Main Menu

, then

MANUAL

and then

select

MAINS VOLTAGE

.

• The

Mains Voltage

test appears on the display, and the test begins immediately.

Display voltages are continuously updated until another test is selected.

baw047f.eps

• Press the

print data

key at any time to generate a printout of the latest

measurement(s)

.

• Press

previous

or

esc/stop

at any time to return to the

MAIN MENU.

Dual Lead Voltage Test

To initiate the Dual Lead Voltage test, select

DUAL LEAD

from the

Mains Voltage

test, or select the Dual Lead Voltage Test from the

MAIN MENU

.

The

Dual Lead Voltage

test appears, and the test begins immediately. Display values

are continuously updated until another test is selected.

baw048f.eps

To take a

Dual Lead Voltage

measurement:

• Apply the voltage to be measured to the RED and BLACK terminals.
• Press the

print data

key at any time to generate a printout of the latest

measurement(s)

.