Fluke Biomedical 601PRO XL User Manual
Page 56

6 0 1 P R O S E R I E S
X L
3 - 1 6
Shortcut Key 0: Mains Voltage Test and Dual Lead Voltage Test
This test can be initiated either from the
test-shortcut
keys or the
MAIN MENU
. The
Mains Voltage Test measures and displays the Mains voltages from L1 to Earth, L2
to Earth, and L1 to L2. The Dual Lead Voltage Test measures the voltage between
the RED and BLACK inputs.
Important:
Power to the 601PRO front panel outlet is off during the Mains Voltage
Test, and on during the Dual Lead Voltage Test.
To measure the mains voltage (L1-EARTH, L2-EARTH, L1-L2), press
shortcut key
0
or press
TESTS/AUTOMODES
from the
Main Menu
, then
MANUAL
and then
select
MAINS VOLTAGE
.
• The
Mains Voltage
test appears on the display, and the test begins immediately.
Display voltages are continuously updated until another test is selected.
baw047f.eps
• Press the
print data
key at any time to generate a printout of the latest
measurement(s)
.
• Press
previous
or
esc/stop
at any time to return to the
MAIN MENU.
Dual Lead Voltage Test
To initiate the Dual Lead Voltage test, select
DUAL LEAD
from the
Mains Voltage
test, or select the Dual Lead Voltage Test from the
MAIN MENU
.
The
Dual Lead Voltage
test appears, and the test begins immediately. Display values
are continuously updated until another test is selected.
baw048f.eps
To take a
Dual Lead Voltage
measurement:
• Apply the voltage to be measured to the RED and BLACK terminals.
• Press the
print data
key at any time to generate a printout of the latest
measurement(s)
.