Fluke Biomedical 601PRO XL User Manual
Page 86

6 0 1 P R O S E R I E S
X L
4 - 6
At the following screen, enter the
Technician Name
and press
ENT
.
baw078f.eps
☛
Note:
If
templates
was selected during setup, then prompts for Serial Number,
Control Number, and Location will also be displayed at this point.
If desired, enter comments to store in the result record as shown below. Press
ENT
to
print test results and return to the
MAIN MENU
.
baw079f.eps
☛
Note:
The maximum number of characters allowed for the
Technician Name
is 15; the maximum for the
Test Comments
is 40.
Warning:
Caution should be taken, as high voltage (110 percent of mains, or 500
VDC) will appear on all leads - particularly when test clips are in place
and only one or two leads are used. Leads should not be left on any
terminal that is not specified for the number of leads selected, as
incorrect readings can occur.
☛
Note:
Press the
Esc
key to discontinue an
Auto/Step
sequence at the
Enter
Technician Name
and
Test Comments
screens.
When performing an
Auto
sequence with startup or power off delays, press the
NEXT
key once the DUT is ready to proceed with the testing; otherwise, the testing of the
DUT will continue once the delay has expired. This is most beneficial for equipment
that has a variable startup or power off time and allows the operator to skip the
remainder of either delay by pressing the
NEXT
key.
baw080f.eps