6 0 1 P R O S E R I E S

X L

4 - 6

At the following screen, enter the

Technician Name

and press

ENT

.

baw078f.eps

Note:

If

templates

was selected during setup, then prompts for Serial Number,

Control Number, and Location will also be displayed at this point.

If desired, enter comments to store in the result record as shown below. Press

ENT

to

print test results and return to the

MAIN MENU

.

baw079f.eps

Note:

The maximum number of characters allowed for the

Technician Name

is 15; the maximum for the

Test Comments

is 40.

Warning:

Caution should be taken, as high voltage (110 percent of mains, or 500

VDC) will appear on all leads - particularly when test clips are in place

and only one or two leads are used. Leads should not be left on any

terminal that is not specified for the number of leads selected, as

incorrect readings can occur.

Note:

Press the

Esc

key to discontinue an

Auto/Step

sequence at the

Enter

Technician Name

and

Test Comments

screens.

When performing an

Auto

sequence with startup or power off delays, press the

NEXT

key once the DUT is ready to proceed with the testing; otherwise, the testing of the

DUT will continue once the delay has expired. This is most beneficial for equipment

that has a variable startup or power off time and allows the operator to skip the

remainder of either delay by pressing the

NEXT

key.

baw080f.eps