Configuring raid network and snmp settings – Grass Valley K2 Storage System Instruction Manual v.3.2 Sep.24 2007 User Manual
Page 161

September 7, 2007
K2 Storage System Instruction Manual
161
Configuring RAID
•
“Prerequisites for initial configuration” on page 155
to confirm that the
RAID storage is ready for configuration.
Configuring RAID network and SNMP settings
Use the Storage Utility to configure the following settings for the Level 3 RAID
controller:
• IP address
• Subnet mask
• Gateway Address
• SNMP trap destinations
For Level 3 RAID, network and SNMP settings are set and stored on the RAID
controller module, so the combined RAID storage devices, including the optional
Expansion chassis, exist as a single entity on the control network.
The RAID storage device is configured by default for the SNMP community name
“public”. If your site’s policies require using a different SNMP community name,
contact your Grass Valley representative.
To configure these settings, do the following:
1. Launch Storage Utility from the K2 System Configuration application as in the
previous procedure.
2. As prompted, wait while Storage Utility gathers system information, then Storage
Utility opens.
3. In Storage Utility tree view, expand the node for the L3 RAID, right-click the icon
for a L3 RAID controller, and select
Configuration | Network Properties
.
The Controller Network Settings dialog box opens.