STC-2000A DEPOSITION CONTROLLER

y

Control Loop Quality Limit

This parameter is used to set a threshold on acceptable control loop quality. When the threshold is

exceeded a sensor fail condition is assumed and the system will process this condition as configured. A
programmed value of 0 disables this test. A value of one gives the tightest limits (5%).
____________________________________________________________________
Parameter

XTL STABILITY

Select Value; Range 0 - 9

Units

None

Crystal Stability Limit

This parameter is used to set a threshold on acceptable sensing crystal performance. When the

threshold is exceeded a sensor fail condition is assumed and the system will process this condition as
configured. A programmed value of 0 disables this test. A value of one gives the tightest limits.
____________________________________________________________________
Parameter

XTAL LIFE BOUNDS

Percentage Value; Range is 0.0 - 100.0

Units

Percent

Crystal Life Limit

This parameter is used to provide a setpoint on the allowable sensing crystal usage. This setpoint

may be utilized through the I/O system as a warning or interlock that sensing crystals need attention.
____________________________________________________________________

The following three parameters are used to select the data type and scale factors for the RunTime

screen graphics area. These parameter values are always reinstated when a film is started.
____________________________________________________________________
Parameter

PLOT VERT SCALE

Vertical

Scale;

Select Specific Values 1/5/10/50/100

Units

Å/s or % Power as selected

Vertical Scale RunTime Display

This parameter sets the vertical scale factor of the graphics display on the RunTime Screen. If

Rate deviation is the type of data being displayed, generally setting this value to 10 will give a good display
of the deposition control.
____________________________________________________________________
Parameter

PLOT HORIZ SCALE

Horizontal Samples; Range is 1-600

Units

Samples Per Display Point

Horizontal Scale RunTime Display

This parameter sets the horizontal sweep rate of the graphics display on the RunTime screen.

Samples are taken at a rate of 4 per second. A programmed value of 1 will display every sample. The
X-axis is 200 samples wide on the RunTime display. By increasing the sample number the sweep rate is
slowed down. Proper selection of this number allows an entire deposition cycle to be displayed on one
sweep of the screen, if desired.
____________________________________________________________________
Parameter

DATA PLOT TYPE

Initial Plot, Select 1 of 3 choices:

Rate

Rate

Deviation

Power

Units

Å/S or % Power as selected

Data Plot Type

This parameter selects the type of information to be displayed in the graphics area of the RunTime

screen.
____________________________________________________________________
Parameter

Source Sensor MAP SELECT

Select Value; Range is 1 – 10

Units

maps

Source Sensor Map Select

SECTION 3.XX

page 82 of 292