STC-2000A DEPOSITION CONTROLLER
y
access: SRCSHUT bitmap true(1) when source shutter open, false(0) if closed. One per active source ident 
(1..8) LSB to MSB. In non-codeposition systems, only one shutter can be open at any one time. 
 
0 ALL SOURCE SHUTTERS CLOSED
1 Source # 1 SRCSHUT OPEN
2 Source # 2 SRCSHUT OPEN
4 Source # 3 SRCSHUT OPEN
8 Source # 4 SRCSHUT OPEN
16 Source # 5 SRCSHUT OPEN
32 Source # 6 SRCSHUT OPEN
64 Source # 7 SRCSHUT OPEN
128 Source # 8 SRCSHUT OPEN 
255 ALL SOURCE SHUTTERS OPENED 
______________________________________________________________________ 
Parameter 9 
Phase Status
Value Range
0 To 17
Units None 
This parameter indicates the phase that the STC-2000A is in according to the following Table: 
 
0 Power
Up 1 Stopped
2 Crystal
Verify
3 Rise
1
4 Soak
1
5 Rise
2
6 Soak
2
7 Soak
Hold
8 Shutter
Delay
9 Deposit
10
Rate Ramp
11
Deposit After Rate Ramp
12
Time Power Recovery
13
Rise 3
14
Idle
15
Idle After Time Power
16 Manual
17 Ready
Table 21.3: Phase Status. 
This is a read only parameter. 
access: SQPHS sequencer phase, holds one of PCD_class (process code) data. 
0 PCD_pup Power Up.
1 PCD_szp Stopped/Zero power.
2 PCD_bxvfy Crystal Verify.
3 PCD_ris1 Rise 1.
4 PCD_sok1 Soak 1.
5 PCD_ris2 Rise 2.
6 PCD_sok2 Soak 2.
7 PCD_skhl Soak Hold.
8 PCD_sdly Shutter Delay.
9 PCD_dpos Deposit.
10 PCD_rtrp Rate Ramp.
11 PCD_dpo2 Deposit phase after a rate ramp
12 PCD_tpwr Time power recovery phase
13 PCD_ris3 Idle ramp.
14 PCD_idlpr Idle power after successful deposit.
15 PCD_idltp Idle after ending deposit with a time/power phase.
16 PCD_manl Manual phase
17 PCD_rdy Ready phase
18 PCD_index Indexing to pocket
19 PCD_fhold Film Holding before rise/soaks.
______________________________________________________________________ 
 
Parameter 10 
Crystal Shutter
Status
Value Range
0,1
Units Open,
Closed
This parameter indicates if the crystal shutter is closed. A returned value of 0 indicate that the shutter is 
open (making measurements) and a value of 1 indicates that it is closed. This is a read only parameter. 
SECTION 6.XX