INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 261
p
STC-2000A DEPOSITION CONTROLLER
y
Command i
Response A0
Description
Memory Module not installed.
______________________________________________________________________
Command: ^
Mnemonic: EXCT_trapinfo
Parameters:
? or ! with no delimiters or U followed by ? or ! for UIFC.
Description: sys_trap
information display (?), clear (!).
Returns:
[RZCD_AOK] if proper parameter value was sent. If was a ^? or ^U? query then returns a result
of the form:
J,N
where J and N are major and minor sys_trap numbers whose meaning is dependent on
the source code. Values of 0,0 indicate no captured error states.
[RZCD_ILDV] if 2nd character (parameter) is other than ? or ! or U ; or if 2nd character is a U
and third character is other than ? or !.
[RZCD_ILCD] if command expression too long or too short (more than correct trailing parameters
or missing any argument parameter info).
Examples:
^?
query current state of SCUP sys_trap error codes. Returns valid reply per syntax above such as:
A1,2
^! clear sys_trap error codes to 0,0 . Armed to get new error conditions. Replies with just the
RZCD_AOK value.
^ missing needed argument, returns RZCD_ILCD.
^ABC More argument than needed, returns RZCD_ILCD.
^4 Argument wrong value, returns RZCD_ILDV.
______________________________________________________________________
Command: j
Mnemonic: EXCT_flush
Parameters: =2001
Description:
Flush non-vol and reboot as if a Service-Purge performed. The syntax of the j command
allows spaces around the '=' operator. The numeric value of 2001 is a security key to
insure that the function is only activated intentionally to purge parameter, sequencing,
process accounting and I/O memory.
Returns:
[RZCD_AOK] if proper parameter value was sent. Unit will reboot after approximately 2 seconds,
and recover to a runtime mode with the system 'hung' in a state which requires user
acknowledgement (via front panel or RS232 event) before deposition can occur.
[RZCD_ILSN] if too many arguments, wrong delimiters, etc. such that format is not of the form
j=NUMBER.
[RZCD_ILDV] if NUMBER following the = operator is other than the correct value of 2001.
______________________________________________________________________
Command: l
Mnemonic: EXCT_CfgManip
SECTION 6.XX
page 261 of 292