Sensor head installation, Stc-2000a deposition controller, Electrical connections and descriptions – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 142: Figure- : 4.1 head mounting dimensions

p
STC-2000A DEPOSITION CONTROLLER
y
Electrical Connections And Descriptions
SECTION 4.0
Sensor Head Installation
Following is a diagram of a typical evaporator installation. This diagram is repeated and described further
in Section 3 (see fig. 3.7).
CRYSTAL SENSOR
SHUTTER
DATA AQUISITION
COMPUTER
CHAMBER
DEPOSITION
SOURCE
OSCILLATOR
Memory Module
CHART RECORDER
WARNING
!
I/O 4 I/O 3 I/O 2 I/O 1
SENSOR
SOURCE
INPUTS / OUTPUTS
1
2
1
2
FUSE: 2 X 2.00 AMP
QUICK-ACTING (F) 250V
90-264 VAC, 50-60 Hz, 230 VA MAX
THE POWER CORD PROTECTIVE
GROUNDING CONDUCTOR MUST
BE CONNECTED TO GROUND, NO
USER SERVICEABLE PARTS
INSIDE. REFER SERVICING TO
QUALIFIED PERSONNEL
instruments
R
S
2
3
2
M
E
M
Sensor Installation
As a general rule the sensor head should not be installed closer than 10 inches to the evaporation
source. This minimum distance will generally provide adequate measurement sensitivity while reducing
the possibility of the source spattering small particles onto the sensor. Even small particles hitting the
crystal surface may cause the crystal to become unstable or stop oscillating completely.
The sensor should be shielded from the evaporant source by a shutter or other means when
evaporant material is being initially conditioned or outgassed.
Water Cooling
Microdot Coaxial Connector
#4-40 Tap -- 4 holes
Equally Spaced
on o0.731 B.C.
Figure- : 4.1 Head Mounting Dimensions.
SECTION 4.XX
page 142 of 292