STC-2000A DEPOSITION CONTROLLER
y
Detailed Film Parameter Descriptions
____________________________________________________________________
Parameter
DENSITY
DENSITY VALUE RANGE IS 0.40 - 99.99
Units
gm/CC
Material Density
The Density parameter refers to the measured material in gm/CC. This constant is normally the
bulk material value but is sometimes different due to deposition and film growth conditions. This value is
utilized in the thickness equation to convert measured mass to a thickness value. See Table 4.1 in Section
4.3 for an extensive value list. See Section 4.2 for calibration information.
____________________________________________________________________
Parameter
Z FACTOR
Z FACTOR VALUE RANGE IS 0.100 - 9.999
Units
None
Material Z-Factor
The Z-Factor parameter refers to the elastic properties of the measured material. This value is
utilized in the thickness equation to match the acoustical properties of the film being measured to the
acoustic properties of the base quartz material of the sensor crystal. This correction is necessary to insure
accurate measurements when sensor crystal shifts of greater than 10% are realized. See Table 4.1 for an
extensive material value list. See Section 4.2 for calibration information.
____________________________________________________________________
Parameter
SETPOINT THK LIM
THICKNESS VALUE RANGE IS 0 - 999.999 KA
Units
Kilo
Angstroms
Setpoint Thickness Limit
The SETPOINT THICKNESS parameter is used to provide a comparison point for the STC-
2000A Thickness Setpoint Event. This event will be triggered whenever the thickness display equals or
exceeds the setpoint value. The final or end thickness of a layer is layer dependent and is determined by
the value programmed in the process, it should not be confused with this parameter which is film
associated.
____________________________________________________________________
Parameter
Final Thickness Limit (Trigger)
non-sequencing only
THICKNESS VALUE RANGE IS 0.000 to 999.999 KÅ
Units
Kilo
Angstroms
Final Thickness Limit
The FINAL THICKNESS LIMIT parameter is only present in the non-sequencing mode. It is used
to provide an end thickness value whose counterpart in the sequencing mode is the thickness value in a
process program. This value determines the end of the deposit for the active film.
____________________________________________________________________
Parameter
SETPOINT TIME LIM
ELAPSED TIME RANGE IS 0:00 - 99:59
Units
Min:Sec
Setpoint Time Limit
The SETPOINT TIME LIMIT parameter is used to provide a comparison point for the STC-
2000A Timer Setpoint Event status. The timer relay status is set when this time is exceeded.
____________________________________________________________________
P
arameter
SOAK 1 PWR VALUE
Percentage Value: Range is 0.0 - 100.0
Units
Percent
Soak 1 Power Value
SECTION 3.XX