beautypg.com

Product programming summary, Summary – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 137

background image

p

STC-2000A DEPOSITION CONTROLLER

y

Product programming summary

SECTION 3.22

Summary


In all of the following cases, use BAKs and/or BYEs for a return path. The fixed STATUS key can be used
at any time or point to return to the RunTime screen. [Screen in the tables refers to the screen currently
displayed.]

To program for PROCESS NON-SEQUENCING...

Screen

Key[s] to press

RunTime fixed

MENU

MAIN MENU

SERVICE

Service menu

SEQ [en]ABLE to NON SEQ + ACCEPT



To program for PROCESS SEQUENCING...

Screen

Key[s] to press

RunTime fixed

MENU

MAIN MENU

SERVICE

Service menu

NON SEQ to SEQ [en]ABLE + ACCEPT



To put the STC-2000A into TEST MODE...

Screen

Key[s] to press

RunTime fixed

MENU

MAIN MENU

SERVICE

Service menu

TEST OFF to TEST ON + ACCEPT

Subsequent power cycling automatically sets TEST MODE to OFF.


To PURGE programmed memory...

Screen

Key[s] to press

RunTime fixed

MENU

MAIN MENU

SERVICE

Service menu

AS IS to PURGE + ACCEPT

Exercise CAUTION PLEASE! See cautions in sections 1.8 and 3.19.


To return to FACTORY settings (programmable elements)...

Screen

Key[s] to press

RunTime fixed

MENU

MAIN MENU

SERVICE

Service menu

AS IS to PURGE to FACTORY + ACCEPT

SECTION 3.XX

page 137 of 292