STC-2000A DEPOSITION CONTROLLER
y
Product programming summary
SECTION 3.22
Summary
In all of the following cases, use BAKs and/or BYEs for a return path. The fixed STATUS key can be used
at any time or point to return to the RunTime screen. [Screen in the tables refers to the screen currently
displayed.]
To program for PROCESS NON-SEQUENCING...
Screen
Key[s] to press
RunTime fixed
MENU
MAIN MENU
SERVICE
Service menu
SEQ [en]ABLE to NON SEQ + ACCEPT
To program for PROCESS SEQUENCING...
Screen
Key[s] to press
RunTime fixed
MENU
MAIN MENU
SERVICE
Service menu
NON SEQ to SEQ [en]ABLE + ACCEPT
To put the STC-2000A into TEST MODE...
Screen
Key[s] to press
RunTime fixed
MENU
MAIN MENU
SERVICE
Service menu
TEST OFF to TEST ON + ACCEPT
Subsequent power cycling automatically sets TEST MODE to OFF.
To PURGE programmed memory...
Screen
Key[s] to press
RunTime fixed
MENU
MAIN MENU
SERVICE
Service menu
AS IS to PURGE + ACCEPT
Exercise CAUTION PLEASE! See cautions in sections 1.8 and 3.19.
To return to FACTORY settings (programmable elements)...
Screen
Key[s] to press
RunTime fixed
MENU
MAIN MENU
SERVICE
Service menu
AS IS to PURGE to FACTORY + ACCEPT
SECTION 3.XX