INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 207
p
STC-2000A DEPOSITION CONTROLLER
y
[RZCD_ILDV] (H or I) if film #, parameter #, or new value is of improper value.
[RZCD_ILSN] if too many arguments, wrong delimiters, etc..
______________________________________________________________________
Film parameter access values:
______________________________________________________________________
Parameter 1
Density
access:
DENS
Value Range
0.40 - 99.99
Units
gm/cc
Reads Or Sets The Density Parameter depending which prefix letter A or B is used.
______________________________________________________________________
Parameter 2
Z Factor
access:
ZRAT
io
Value Range
0.100 - 9.999
Units None
Reads Or Sets The Z Factor Parameter depending which prefix letter A or B is used.
______________________________________________________________________
(Obsolete)
Parameter 3
Tooling Xtal 1
access:
OBSOLETE Was sensor 1 tool
Value Range
10.0 - 400.0
Units
Percent
______________________________________________________________________
(Obsolete)
Parameter 4
Tooling Xtal 2
access:
OBSOLETE Was sensor 2 tool
Value Range
10.0 - 400.0
Units
Percent
Reads or sets depending which prefix letter A or B is used, the tooling factor parameters.
These correspond to the two sensors available on the STC-2000A.
______________________________________________________________________
(Obsolete)
Parameter 5
Start Xtal #
access:
OBSOLETE Was initial channel
Value Range
0,1
Units
0 is Sensor 1, 1 is Sensor 2
This indicates which of the 2 sensor crystals will be used for a deposition run.
______________________________________________________________________
Parameter 6
Setpoint Thk Lim
access:
STHK
Setpoint Thickness Limit
Value Range
0.000 - 999.999
Units
Kilo
Angstroms
The setpoint thickness parameter is used to provide a comparison point for the STC-
2000A thickness setpoint event. This event will be triggered whenever the thickness
display equals or exceeds the setpoint value.
______________________________________________________________________
(non-sequencing
mode
only)
Parameter 7
Final Thickness Limit (Trigger)
access:
FTHK final thickness, end of run trigger
Value Range
0.000 - 999.999
Units
Kilo
Angstroms
SECTION 6.XX
page 207 of 292