Read runtime data group 4 -- command e – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 219
p
STC-2000A DEPOSITION CONTROLLER
y
Units
None
______________________________________________________________________
Parameter 31
I/O Slot 4 Type
access value:
REQIO4 Require Input/Output 4 in system, CFGIOS_class
Value Range
UNUS
ED
(DISABLED) / INPUT / OUTPUT
Units
None
______________________________________________________________________
Parameter 32
Memory Module IFC
access value:
REQMEMMOD Require memory module option installation. Not auto detectable Boolean 1/0 True/False.
Value Range
OFF/ON
Units
None
______________________________________________________________________
______________________________________________________________________
Read, Write System Parameters Command Examples
Command
C10 or C 10
Response A15:33
Description
Read system parameter # 10, system clock time. Result is 15:33 hrs:mins, 3:33 p.m. time.
Command C18
Response A5
Description
Read system parameter # 18, NA.
Command
D11 = 70689
Response A
Description
Set system parameter 11, calendar date to July 6, 1989. Command received and accepted.
Command
D 14= 1
Response
A (Possible missed due baud change)
Description
Set baud rate 1200 baud.
Table 6.5: Group 3 Examples.
______________________________________________________________________
______________________________________________________________________
Read RunTime Data
Read RunTime Data
Group 4 --
Command E
(E)
Read RunTime Data. These commands are used for reading back real time numerical information
from STC-2000A. This includes information such as rate, thickness, power, etc. The group
command is the letter E followed by a parameter number.
Command: E
Mnemonic:
EXCT_rdata (read runtime data)
Parameters:
[1..61] corresponding to real time data types PWRLVL..XSTAB8 .
Description:
Requests a read back of real time (not fixed parameters) system data. See the list of
data and the corresponding ident code below.
Returns:
SECTION 6.XX
page 219 of 292