STC-2000A DEPOSITION CONTROLLER

y

Shutter Delay Timeout

This parameter sets a time limit for the delayed shutter operation to be completed. The time period

begins at the beginning of the shutter delay cycle. The sensor is exposed to evaporant with the substrates
remaining shielded. When rate control is established to the programmed accuracy, three things happen.
The substrate shutter is opened, accumulated thickness set to zero, and the programmed deposition profile
is then started. If the specified rate control accuracy cannot be met within this time limit, the deposition will
be aborted.
____________________________________________________________________
Parameter

SHTR DLY QUALITY

Range is 1 - 50

Units

Percent


Shutter Delay Quality

This parameter sets the rate control accuracy that must be established for a period of five seconds

in order to complete the delayed shutter sequence. The maximum accuracy settable is 1 percent of the
desired setpoint. However, this is internally limited to 1 angstrom per second for noise and resolution
restrictions.
____________________________________________________________________

Rate Sampling (has 4 dependent parameters)

This operating mode, when enabled, allows the user to expose the crystal sensor for short periods

of time to the evaporant stream, establish rate control, and then shutter the sensor maintaining the power
setting established during exposure. For relatively stable deposition sources or systems that must maintain
vacuum integrity for long periods of time, crystal sensor life can be greatly extended. Two sampling modes
are available. One has the exposure time strictly time controlled, the other will expose the sensor only long
enough to establish the required accuracy. The exposure period for either mode is also triggerable
externally.

Sample and Hold

Figure- 3.11: Rate Sampling, Sample and Hold Operation.


Parameter

RATE SAMPLING

Rate Sample choices (1 of 3):

OFF
TIMED

MODE

INTELLIGENT

MODE

Units

None

Enabling Rate Sampling

This switch type parameter will allow the rate sampling parameters to become active and the mode

functional on an individual film basis. If set to Off the parameter window cursor will not access the
associated parameters. If Timed Mode is selected, the crystal sensor will be exposed to evaporant for the
time specified by the sample dwell parameter, and will sample at the interval specified by the sample
interval parameter. If Intelligent Mode is selected, then the duration of the sample will only be long enough
to establish the desired accuracy. [The following 4 parameters are dependent upon this parameter.]
____________________________________________________________________
Parameter

RATE SAMPLE

On/Off Value: Off or On

Units

On/Off


Enabling Rate Sampling

SECTION 3.XX

page 80 of 292