STC-2000A DEPOSITION CONTROLLER
y
This parameter returns the current deviation from the programmed rate setpoint. A negative number
indicates the actual value is below the programmed one. A returned number of -123 indicates that the rate
is 12.3 Å/s below whatever the programmed value is.
______________________________________________________________________
Parameter 7
Crystal Life (primary sensor)
access index:
XLIFE crystal life value, 0..99 percent., primary crystal.
Value Range
0 To 99
Units
Percent
This parameter returns the percentage of the crystal life that is left on the primary crystal sensor. It is the
same number as on the RunTime display.
______________________________________________________________________
Parameter 8
Sensor Channel And Status (active channel)
access index:
XCHNL crystal channel in use.
Value Range
0 To 5
Units
None
This is a composite number where the lsb is 0 for sensor 1 selected, 1 for sensor 2 selected; bit 1 is 0 if the
sensor is not good, 1 if it is good; and bit 2 is 0 if the sensor has not failed, and 1 if the sensor has failed. If
bits 1 and 2 are both 0, this indicates that the sensor status is unknown.
______________________________________________________________________
Parameter 9
Crystal Stability Value
access index:
XSTAB crystal stability value
Value Range
0 To 99
Units
None
The tens digit of this number is incremented when the stability level is violated. The units digit indicates
10 x of range violated to get to the next stability level.
______________________________________________________________________
Parameter 10
Process Quality Value
access index:
PQUAL process quality value
Value Range
0 To 99
Units
None
The tens digit of this number is incremented when the quality level is violated. The units digit indicates 10
x of range violated to get to the next quality level.
______________________________________________________________________
Parameter 11
Control Quality Value
access index:
CQUAL control quality, instantaneous
Value Range
0 To 99
Units
Percent
This parameter indicates how good the control loop is. It is the deviation expressed as a percentage of the
setpoint.
______________________________________________________________________
Parameter 12
Sample Hold Time
access index:
SHTIME sample hold time
Value Range
0 To 5999
Units
Seconds
This is the time remaining in the current sample/hold phase.
______________________________________________________________________
Parameter 13
Crystal Frequency
access index:
FREQ Crystal Frequency 5 Mhz to 6 Mhz to .1 hertz, 0 if bad. from primary (when not
averaging) sensor crystal.
SECTION 6.XX
page 221 of 292