INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 213

p
STC-2000A DEPOSITION CONTROLLER
y
access:
CLQU control loop quality limit
Value Range
0 - 9
Units
None
This parameter is used to set a threshold on acceptable control loop quality. When the
threshold is exceeded a sensor fail condition is assumed and the system will process this
condition as configured. A programmed value of 0 disables this test. A value of one
gives the tightest limits (5%).
______________________________________________________________________
Parameter 39
Xtal Stability
access:
XLQU crystal oscillation quality limit
Value Range
0 - 9
Units
None
This parameter is used to set a threshold on acceptable sensing crystal performance.
When the threshold is exceeded a sensor fail condition is assumed and the system will
process this condition as configured. A programmed value of 0 disables this test. A value
of one gives the tightest limits.
______________________________________________________________________
Parameter 40
Xtal Life Bounds
access:
LIFB life bounds for I/O event
Value Range
0.0 - 100.0
Units
Percent
This parameter is used to provide a setpoint on the allowable sensing crystal usage. This
setpoint may be utilized through the I/O system as a warning or interlock that sensing
crystals need attention.
______________________________________________________________________
Parameter 41
Plot Vertical Scale
access:
VSCL vertical scale units, 1..5..10..etc.
Value Range
0-4
Units
1,5,10,50,100 Å/s Or % Power As Selected
This parameter sets the vertical scale factor of the graphics display on the RunTime
screen.
______________________________________________________________________
Parameter 42
Plot Horizontal Scale
access:
HSPL horizontal sampling, 1,2,3.. samples per point
Value Range
1-600
Units
Samples Per Display Point
This parameter sets the horizontal sweep rate of the graphics display on the RunTime
screen. Samples are taken at a rate of 4 per second. A programmed value of 1 will display
every sample. The x-axis is 200 samples wide on the RunTime display.
______________________________________________________________________
Parameter 43
Data Plot Type
access:
IPLT initial plot variable, chosen at start time
Value Range
0-2
Units
0 is Rate, 1 is Rate Deviation, 2 is Power
This parameter selects the type of information to be displayed on the RunTime graphics
screen.
______________________________________________________________________
Parameter 44
Source Sensor MAP SELECT
access:
SMAP Source/Sensor map associated, 1..30
Value Range
1 - 30
SECTION 6.XX
page 213 of 292