Reset power lost status – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 205

p
STC-2000A DEPOSITION CONTROLLER
y
@10 = response from SCUP UIFC vrsn string info.
@11 = UIFC Code signature.
@12 = UIFC Bootloader signature.
@13 = UIFC serial/build.
@14 = Base SCUP version for UIFC.
@15 = UIFC Glossary covenant.
@16 = UIFC sys_trap codes. In nnnn,nnnn form.
Read at power up, can be refreshed for recent reading from UIFC via ^U? 
command. Reset from SCUP via ^U! command. Note, the ^U? command 
requests a new sys_trap code from UIFC, but requires a separate @16 to 
interrogate the reading if changed. Once a sys_trap code is non-zero, it will not 
change or be overwritten until cleared from the UIFC front panel or via the YH 
command passthru or the ^U! command. 
@17..@29 reserved for future UIFC codes.
@30 = ReadBack SCUP Memory Module status. Returns CRDSTAT_class integer.
0 CRDSTAT_uninst Uninstalled option.
1 CRDSTAT_error Unavailable or unpresent when spoken to.
2 CRDSTAT_unfmtRO Present but format not recognized.
3 CRDSTAT_unfmt Present but format not recognized.
4 CRDSTAT_wrprot Present but write-protected.
5 CRDSTAT_readwrite Totally available.
@31 = Readback MemModule ProdID & Serial# returns integer, integer where first
integer is 9/10/11 for FRAM memory module w/1/2/3 chips. If chip's status is 
not above unformatted, then returns 0,0. Serial number in range of 1..1000 is one 
which was formatted automatically on SCUP system. Factory registered ones 
start at serial # 2001 and up. This leaves an additional area from 1001..2000 for 
future special purpose. 
@32 = Readback Data Compatibility code returns number,number where 1st number
is product code (such as COBRASCUP #4) and second number is a dialect/data 
representation covenant code which determines whether the data saved to file is 
expected to be interpreted for read/write of params/accounting/sequences/ and 
RLLE steps. If chip's status is not above unfmt, then returns 0,0 . This 
information pertains to the portable data on the CARD. 
@33 = Readback Data Compatibility code returns number,number where 1st number
is product code (such as COBRASCUP #4) and second number is a dialect/data 
representation covenant code which determines whether the data saved to file is 
expected to be interpreted. This information pertains to the SCUP firmware's 
desired/expected information profile. It must MATCH EXACTLY the card in 
order to load/clone. 
______________________________________________________________________ 
Initialization Commands Example 
Command @ 
Response ASTC200/B15 
Description 
The version request message. Where the 'A' might be a 'B' if the instrument was reset 
since the last power loss acknowledgment message. 
______________________________________________________________________
Reset Power Lost Status
This is used to reset the power lost status of the STC-2000A. See Section 3 for a complete description. 
 
Command: ? 
Mnemonic: EXCT_ackpl (acknowledge power loss) 
Parameters: None 
Description: Acknowledge Power Loss status. 
Returns: 
[RZCD_AOK] (an A or B)
SECTION 6.XX
page 205 of 292