Stc-2000a deposition controller – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual
Page 17

p
STC-2000A DEPOSITION CONTROLLER
y
POCKET SELECT
0 – 63 Pockets
ETCHING MODE
OFF/ON
Process Step elements:
[Menu end point: Review Processes
Menu path: Main/Review Processes]
Review Processes (Edit Process)
(Edit Process Steps [1-99]) Mode (Skip, Stop, Auto,
End, Wait), Film# (1-50), Thickness (0.0 - 999.999 KÅ)
Main Menu values:
[Menu end point: Main Menu
Menu path: fixed front panel MENU key/ Main Menu]
Next Active Process (Select)*
1 of 9 processes [sequencing mode only]
Next Active Film (Select)*
1 of 50 films [non-sequencing mode only]
[*NOTE: press review ______ touch key, then press GO2 ____ to select any one of the 9 processes, 10 maps or 50 films]
RunTime Screen parameters:
[Menu end point: RunTime screen
Menu path: programming Menu/ fixed front panel STATUS key/RunTime screen]
(Crystal Quality Indicator Select) L/Q, Loop x, Qual xx
(Crystal Sample Select)
(not selected [blank]), SMPL, SMPL+Time, HOLD+Time
START key sequence
(fixed front panel)
(Note PROCess X: message change after each START process key press)
2
STOP key
(fixed front panel)
Stops the running process
(non-seq mode: stops the 1 inherent process [film])
MANUAL key
(fixed front panel)
Switches to manual mode from a running process only. Toggle action.
ZERO key
(fixed front panel)
Zeroes the thickness value
MENU key
(fixed front panel)
Switches from RunTime screen to MAIN menu screen
STATUS key
(fixed front panel)
Switches to RunTime screen from any menu screen, to large character Run
Time, detailed power and crystal info screens from RunTime screen
Zero Power
[non-active process]
zero channel 1-2, zero all [2
nd
STATUS key press from RunTime]
Zero Thickness
1
zero channel 1-2, zero all, re-verify
[3
rd
STATUS key press from RunTime]
Force Fail
[non-active process]
fail channel 1-2, fail all, re-verify
[3
rd
STATUS key press from RunTime]
1
Zero thickness does not affect film thickness value, only Source Sensor card value (use to set tooling factor, diagnose
problems, etc. )
2
If the PROCess X: status message does not change when the START key is pressed, check for "STOP: INValid XXX "
message or check the OPT/INF menu, page 2 for cards not installed which are enabled in software (MIA). Cycle AC power
OFF
ON to re-sync.
System Configuration values:
[Menu end point: System Config
Menu path: Main/Executive menu/System Configuration]
LCD Contrast / Bias
LOW, MEDIUM, HIGH
Password Lock #
0 - 9999
(Process) Run Number
0 to 9999
Recorder Function
Absolute Rate, Rate Deviation, Power, Thickness, Computer Remote, I/O Control, Off
Recorder Out Channel
1 – 2 selects an analog output channel for use
(if not off and not used as a source)
Real Clock Time
HH/MM
Real Clock Date
MM/DD/YY
I/O Slot 1 Type
UNUSED (DISABLED) / INPUT / OUTPUT
I/O Slot 2 Type
UNUSED (DISABLED) / INPUT / OUTPUT
I/O Slot 3 Type
UNUSED (DISABLED) / INPUT / OUTPUT
I/O Slot 4 Type
UNUSED (DISABLED) / INPUT / OUTPUT
Memory Module IFC
OFF/ON
Communication values:
[Menu end point: Comm. Setup
Menu path: Main/Executive menu/Communications Setup]
COM/IO Lock Code
0 - 9999
Keyboard Beep
OFF / ON
RS232 Baud Rate
300, 1200, 2400, 9600
RS232
Protocol
Sycon,
ASCII
Source Sensor Map elements:
[Menu end point: Review SS Map
Menu path: Main/Review SS Map]
Source Sensor Full Power Volts
2.5, 5, 10 full scale volts
SECTION 1.XX
page 17 of 292