Stc-2000a deposition controller – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 149

background image

p

STC-2000A DEPOSITION CONTROLLER

y

CHART RECORDING OF 3500Å FILM

ANALOG OUTPUT

FOR THICKNESS

3500Е

3000Е

2000Е

1000Е

V
O
L
T
S

1

2

3

4

5

6

7

8

9

10

Figure- 4.4: Example of Thickness Mode Recorder Output.


Rate Mode - The analog recorder output for the rate mode is scaled for a full scale (999.9 Å/s)

reading of 10 volts. Readings up to 100.0 Å/s will be output with a resolution of 0.1 Angstrom and values
above 100.0 Angstroms will be output with a resolution of 1.0 Å/s. This method allows a continuous
recorder accuracy of at least 0.3% and also provides a unique output voltage for any rate value within the
displayable range of values. This output is unipolar with negative rates clamped at zero volts.

Rate Deviation Mode - linear from 0 to

±999.9 A°/S and 0 to 10 Volts. 0 to 999.9 A°/S generated

by voltages between (and including) 5 and 10 volts. 0 to -999.9 A

°/S generated by voltages between (and

including) 5 and 0 volts. In other words, zero rate deviation is midscale at 5 volts.

Power Mode - The analog recorder output for the power mode is scaled for a full-scale (100%)

reading of 10 volts.

RATE - ÅNGSTROMS / second

ANALOG OUPUT

FOR RATE

Slope 5.555 mv per Å/s

between 100 Å/s and 1000 Å/s

Slope 50 mv per Å/s

between 0 and 100 Å/s

100

200

300

400

500

600

700

800

900

1000

V
O
L
T
S

1

2

3

4

5

6

7

8

9

10

Figure- 4.5: Example of Rate Mode Recorder Output.



SECTION 4.XX

page 149 of 292