STC-2000A DEPOSITION CONTROLLER

y

SECTION 3.18

Process Accounting


The process accounting screen displays a summary of the vital parameters of the last four previous

deposition films. It is useful to review this screen at the completion of a deposition run and compare the
results to the previous similar runs. Trends which show changes in source conditions, deposition material
supply or pending crystal failure can sometimes be determined by the comparison of the vital parameters
shown on the process accounting screen.

All the data displayed on this screen can also be 'read' through the computer interface for longer-

term storage (more than four runs) or hard copy data logging. Starting at the Runtime screen...

Press the fixed front panel Menu key to invoke

On the Executive menu, press the Process

the

main

menu.

Accounting

key.

The following is the initial screen when process accounting is selected (page 1 of 4). The first line of the
list is RUN# and it remains stationary as the list is scrolled in order to maintain a fixed frame of reference.
The double arrow down moves the list a section at a time so that the list is displayed in 5 line sections in 4
pages (a 5 line window moving down the list). The double up arrow moves the viewed section of the list in
the same way but the direction is up the list. The scrolling function will wrap. After the 4

th

page, the view

will revert to the 1

st

page when the double down arrow key is pressed. After the 1

st

page, the view will

revert to the 4

th

page when the double up arrow key is pressed.


The following screens show the progression through the list using the double down arrows.

Page 2 of 4

Page 3 of 4


Page 4 of 4

Figure- 3.16: Process Accounting Screens


Please note screens 4 of 4. The XTAL % life is reported for the start condition in the first 3 characters (e.g.
A17, F00, O00, O..) and for the ending condition in the next 3 characters (e.g. A17, F00, O00, O..). The
first of the 3 characters is a letter code that is explained in the table below and the next 2 characters are a
numeric value, which is the XTAL % life. The O letter code has an additional differentiation in that if it is
followed by two periods, this means that there is no sensor card installed for this crystal channel, otherwise

SECTION 3.XX

page 119 of 292