1.8 Functions of the 4-CH, Medium-Speed Strain

Input Module

This module allows up to four inputs of measurement from strain gauges and strain
gauge type sensors at a minimum measurement interval of 100 ms.

-B12, -B35

-NDI

Input terminal

(clamp terminal)

Terminal cover

NDIS terminal

Measurement Input Types

Measurement Input Type

Notation

No measurement

SKIP

Strain

STR

Measurement Range

Strain Input

Measurement Range Type

Notation

Rated Measurement Range

2000 µSTR

2000 uSTR

–2000.0 to 2000.0 µSTR

20000 µSTR

20000 uSTR

–20000 to 20000 µSTR

200000 µSTR

200000 uSTR

–200000 to 200000 µSTR

Measurement Interval, Integration Time, and Filter

You can select from the following measurement intervals for this module.
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 60 s
The integral time and types of filters applied vary depending on the measurement
interval.

For information on filters, see section 2.9, “Measures Against Noise on the MW100 Data

Acquisition Unit.”

Measurement Synchronization

Since measurement is sequential, measurement on each channel is not synchronized.