4-6
IM MW100-01E
Execution Errors
The code is divided into two parts which are displayed alternately on the 7-segment LED;
in the first part, the letter E appears in the left digit with the hundreds digit of the error
code to the right, and the second part consists of the last two digits of the error code.
Display
Possible Problem
Corrective Action
E301
CF card error detected.
Do not eject or otherwise disturb the card while being accessed.
E302
No enough free space on CF card.
Delete unneeded files to free up space. Replace the CF card.
E303
CF card is write-protected.
Check write permissions.
E311
CF card not inserted.
Insert the CF card correctly.
E312
CF card format damaged.
Check the CF card. Please format the CF card.
E313
CF card damaged or not formatted.
The file may be damaged. Format or replace the CF card.
E314
File is write-protected.
Check write permissions.
E315
No such file or directory.
Check the files and folders.*
E316
Number of files exceeds the upper limit.
Delete unneeded files to reduce the number of files.
E317
Invalid file or directory name.
Check the files and folders.*
E318
Unknown file type.
Check the files.
E319
Same name of file or directory already exists.
Check the files and folders.*
E320
Invalid file or directory operation.
Check the files and folders.*
E321
File is in use.
Wait until access is finished.
E331
Setting file not found.
Check the name of the setting file.
E332
Setting file is broken.
Could not load setting file because it is corrupted.
E341
FIFO buffer overflow.
You must reduce the time required to store files.
Delete unneeded files to free up space.
E342
Data to be saved to file not found.
Check the settings.
E343
Power failed while opening file.
Files may have been damaged.
Take appropriate action for power failure.
E344
Some or all data prior to power outage could not Do not change the CF card during a power failure.
be recovered.
E345
Could not restart recording after recovery from
Perform the record start operation.
power outage.
E346
Recording could not be started due to power
Perform the re-recording start operation.
outage.
Take appropriate action for power failure.
* May occur in the MW100 internal processing (during an abnormality)
Communication Command Errors
The code is divided into two parts which are displayed alternately on the 7-segment LED;
in the first part, the letter E appears in the left digit with the hundreds digit of the error
code to the right, and the second part consists of the last two digits of the error code.
Display
Possible Problem
Corrective Action
E401
Command string too long.
Keep the command within 2047 Bytes from first character to
terminator.
E402
Too many commands enumerated.
Set the number of enumerated commands within 99.
E403
Invalid type of commands enumerated.
Send the commands without enumerating them.
E404
Invalid command.
Confirm the command name.
E405
Not allowed to execute this command.
Login at a level that allows execution of this command.
E406
Cannot execute due to different operation mode. Switch to a mode that allows execution of this command.
E407
Invalid number of parameters.
Check the number of parameters.
E408
Parameter string too long.
Keep the length of individual parameters within 512 Bytes.
E411
Daylight saving time function not available.
Check the daylight saving time settings.
Not available with the current model.
E412
Temperature unit selection not available.
Not available with the current model.
E413
MATH option not available.
Not available with the current model.
E414
Serial communication interface option not available. Not available with the current model.
E415
Report option not available.
Not available with the current model.
4.1 Error Display on the 7-Segment LED and Corrective Actions