5-40

IM MW100-01E

Pt100

*1

JPt100

*1

Pt100 (high res.)

JPt100 (high res.)

Ni100 SAMA

*2

Ni100 DIN

*2

Ni120

*3

Pt50

Cu10 GE

*4

Cu10 L&N

*4

Cu10 WEED

*4

Cu10 BAILEY

*4

J263B

Cu10 at 20°C

alpha=0.00392

Cu10 at 20°C

alpha=0.00393

Cu25 at 0°C

alpha=0.00425

Cu53 at 0°C

alpha=

0.00426035

Cu100 at 0°C

alpha=0.00425

Pt25

(JPt100/4)

Cu10 GE

(high resolution)

Cu10 L&N

(high resolution)

Cu10 WEED

(high resolution)

Cu10 BAILEY

(high resolution)
PT100GOST

Cu100GOST

Cu50GOST

Cu10GOST

Level

Contact input

–200.0 to 600.0°C

–200.0 to 550.0°C

–140.00 to 150.00°C

–140.00 to 150.00°C

–200.0 to 250.0°C

–60.0 to 180.0°C

–70.0 to 200.0°C

–200.0 to 550.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

0.0 to 300.0K
–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

–50.0 to 150.0°C

–50.0 to 150.0°C
–200.0 to 550.0°C
–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 300.0°C

–200.0 to 600.0°C

–200.0 to 200.0°C

–200.0 to 200.0°C

–200.0 to 200.0°C

Vth = 2.4 V

±(0.05% of rdg + 0.3°C)

±(0.05% of rdg + 0.3°C)

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 2°C)

±(0.05% of rdg + 0.3K)

±(0.1% of rdg + 2°C)

±(0.1% of rdg + 0.5°C)

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 0.5°C)

±(0.1% of rdg + 2°C)

±(0.05% of rdg + 0.3°C)

±(0.05% of rdg + 0.3°C)

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 2°C)

Threshold level accuracy: ±0.1 V

ON for 100 Ω or less, OFF for 10 kΩ or more

*5

±(0.1% of rdg + 1.5°C)

±(0.1% of rdg + 1.5°C)

±(0.1% of rdg + 1.5°C)

±(0.2% of rdg + 5°C)

±(0.1% of rdg + 1.5K)

±(0.2% of rdg + 5°C)

±(0.2% of rdg + 2°C)

±(0.1% of rdg + 1.5°C)

±(0.2% of rdg + 2°C)

±(0.2% of rdg + 5°C)

±(0.1% of rdg + 1.5°C)

±(0.1% of rdg + 1.5°C)

±(0.1% of rdg + 1.5°C)

±(0.2% of rdg + 5°C)

Input

Measurement Accuracy

Integration Time: 16.67 ms

or More

Rated

Measurement

Range

Measurement Accuracy

Integration Time: 1.67 ms

Measurement

Range

Type

*1 Pt50: JIS C 1604-1989, JIS C 1606-1989/Pt100: JIS C 1604-1997, IEC 751, DIN IEC 751/JPt100: JIS C 1604-1989, JIS C 1606-1989

*2 SAMA/DIN

*3 McGRAW EDISON COMPANY

*4 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/

Cu10 BAILEY: –20.0 to 250.0°C

*5 Measured using a measurement current of approximately 10 μA at 200 mV range. Threshold level is approximately 0.1 V.

Highest

Resolution

(1 digit)

0.1°C

0.01°C

0.1°C

0.1°C

0.1K

0.1°C

0.1°C

0.1°C

0.1°C

0.1°C

0.1°C

0.1°C

0.1°C

0.1°C

RTD

(Measurement

current: 1 mA)

DI

Measurement interval, integration time, and filter:

Measurement

Interval

100 ms

16.67 ms

20 ms

Auto

*

3

Integration

Time

Filter

Rejected Noise and Notes

500 ms

Rectangular

1 s

2 s

5 s

10, 20, 30, 60 s

36.67 ms

100 ms

*

4

200 ms

*

5

200 ms

Trapezoidal

Rectangular
Cos

600 Hz and its integer multiples

*

2

60 Hz and its integer multiples
50 Hz and its integer multiples

Automatically detects the power supply frequency and set

16.67 or 20 ms

50 Hz or 60 Hz and their integer multiples

10 Hz and its integer multiples
Fc = 5-Hz low-pass filter

Burnout

Detection

Cycle

200 ms

1 s

*

1

1.67 ms

Measurement

interval

*1 When the measurement interval is 100 ms, burnout is detected in one channel per measurement interval.

Therefore, if measurement is started in a burnout condition or after a burnout occurs, burnout cannot be

detected for up to 10 measurements (approximately 1 second).

*2 Because the power supply frequency noise is not rejected, measured values may fluctuate particularly for

temperature measurements using thermocouples. If this happens, make the measurement interval longer, or

use the 4-CH High-Speed Universal Input Module.

*3 For DC power, set to 20 ms.

*4 When synchronizing time by SNTP, the integral time is set to 36.67 ms. Also in this case, noise of 50 Hz, 60

Hz, and their integer multiples is rejected.

*5 When synchronizing time by SNTP, the integral time is set to 100 ms. Also in this case, noise of 10 Hz and its

integer multiples is rejected.

Reference junction compensation:

Switch external/internal by channel, includes remote RJC function