App index – Yokogawa Standalone MW100 User Manual
Page 254
5-41
IM MW100-01E
Specification
1
2
3
4
5
App
Index
Reference junction compensation accuracy:
When measuring temperature greater than or equal to 0°C and
when the temperature of the input terminal is balanced
Type R, S, W: ±1°C
Type K, J, E, T, N, L, U, XK GOST: ±0.5°C
Type N (AWG14), PLATINEL, NiNiMo, WRe3-25,
W/WRe26: ±1°C
Note: Type B and PR40-20 internal RJC is fixed at 0°C
Maximum input voltage: DC voltage at 1-V range or less, TC, RTC, and DI (contact):
±10 VDC (continuous)
Other measurement ranges: ±120 VDC (continuous)
Normal-mode voltage: DC voltage, TC, DI (LEVEL): 1.2 times the range rating or less
(50/60 Hz, peak value including the signal component)
RTD 100 Ω: 50 mVpeak
RTD 10 Ω, 25 Ω, 50 Ω: 10 mVpeak
Normal-mode rejection ratio:
For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ± 0.1%)
For integration time of 1.67 ms: 50/60 Hz is not rejected.
The RTD and resistance ranges indicate the voltage conversion
value when current flows.
Common-mode voltage:600 VACrms (50/60 Hz), reinforced (double) insulation
Common mode rejection ratio (50/60 Hz ±0.1%, 500 Ω unbalanced between minus
measurement terminal and ground)
The RTD and resistance ranges indicate the voltage conversion
value when current flows.
When the integration time is 16.67 ms or more: 120 dB or more
When the integration time is 1.67 ms: 80 dB or more
Common-mode voltage between channels: 120 VACrms (50/60 Hz)
Noise rejection:
Rejection by the integrating A/D converter and the use of low pass
filters
Input resistance:
For DC voltage 1 V range or less and thermocouple range: 10
MΩ or less
For DC voltage 2 V range or higher: Approx. 1 MΩ
Insulation resistance: Between input and ground: 20 MΩ or more (500 VDC)
Input bias current:
10 nA or less (except for burnout setting)
Withstand Voltage:
1000 VAC (50/60 Hz) for one minute between input terminals
3700 VAC (50/60 Hz) for one minute between the input terminal
and earth terminal
Input signal source resistance:
DC voltage, thermocouple: 2 kΩ or less
RTD 50 Ω, 100 Ω types: 10 Ω per line or less
RTD 10 Ω, 25 Ω types: 1 Ω per line or less
Thermocouple burnout: Detection at a specified detection interval per measurement
interval and detection within the thermocouple range (detection
ON/OFF possible)
Up/Down setting is possible
2 kΩ or less normal, 200 kΩ or more disconnected
(shunt capacitance: 0.01 µF or less), detection current approx. 10
µA, detection time approx. 2 ms
Parallel capacity during RTD: 0.01 µF or less
Power consumption: Approx. 1.2 W
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal
cover)
Weight:
Approx. 0.5 kg.
Terminal type:
Clamp, terminal board is detachable
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)
5.5 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10) Specifications